Configuration
56
The firmware updates for supply module and power module may also contain necessary
updates for the remote control. Accordingly, after every firmware update, you need to
check whether the remote control firmware also needs to be updated. Proceed as
follows:
Step Action
1
Make sure that the remote control is activated (see page 48) and switched on.
2
Open the user interface locally at the supply module.
3
Open the page
SFC250-RC
.
4
Check whether the
Status
line shows the message “
SFC250-RC found.
Please update firmware!
”
If the message is displayed, proceed with the next step to install the update.
If however no update of the remote control is required, you can ignore the
following steps.
5
Open the menu item
Firmware Update
on the user interface.
6
Move at least 2 metres away from the supply module and press the button
on the remote control.
Result:
The update starts and the progress is shown on the user interface.
After the update is successfully completed, the message
SFC250-RC active
is
shown in the
Status
line.
The remote control now once again has a firmware that is compatible with the
supply module.
Updating the remote