background image

TPT420

Two pole  
voltage tester

User guide

Test Equipment Depot - 800.517.8431 - 

5 Commonwealth Ave, MA 01801 

 TestEquipmentDepot.com 

Summary of Contents for TPT420

Page 1: ...TPT420 Two pole voltage tester User guide Test Equipment Depot 800 517 8431 5 Commonwealth Ave MA 01801 TestEquipmentDepot com...

Page 2: ...by connecting to a known voltage source e g a suitable 120 230 V outlet or a suitable proving unit before and after use Tests for voltage should be made with both poles Extreme care must be taken whe...

Page 3: ...age is present When there is no indication of voltage present it is still recommended to earth equipment before commencing work A voltage detector of relatively high internal impedance compared to the...

Page 4: ...rotected by an RCD RCBO and Safety Breaker the TPT420 will not trip these devices The phase rotation indication function is simplified which avoids the crossing of the test probes adopted by some 2 po...

Page 5: ...an also be switched on with the torch light button Auto power off The tester is automatically powered off after 5 sec when there is no signal detected by the probes The torch light switches off after...

Page 6: ...age warning LED illuminates 50 VAC 120 VDC and the LCD display is blank then check the batteries This feature must not be used as a means of continuing voltage indications If the low battery indicator...

Page 7: ...ED indicates DC DC During a voltage test L or R LEDs may illuminate For voltages over 1000 V AC and 1500 V DC the LCD will display OL Single pole phase test Hold the tester securely in your hand Conne...

Page 8: ...t good enough Torch light Pressing the torch light button will turn on the light and after approximately 10s it will turn itself off Battery Replacement Remove the probes from any testing point when o...

Page 9: ...r of Electrical and Electronic equipment The registration no is WEE DJ2235XR Users of Megger products in the UK may dispose of them at the end of their useful life by contacting B2B Compliance at www...

Page 10: ...ange indication OL Peak current Is 3 5 mA at 1000 V Measurement Duty 30 s ON operation time 240 s OFF recovery time Internal battery consumption Approx 80 mA Single pole phase test voltage range 100 1...

Page 11: ...a Returns Authorisation number must first be obtained by contacting one of the addresses shown You will be asked to provide key information such as the instrument serial number and fault reported when...

Page 12: ...s user guide is in compliance with Directive 2014 53 EU Other equipment manufactured by Megger Instruments Limited described in this user guide is in compliance with Directives 2014 30 EU and 2014 35...

Page 13: ...M Test Equipment Depot 800 517 8431 5 Commonwealth Ave MA 01801 TestEquipmentDepot com...

Reviews: