2

References marked on 

tester or in instruction 

manual

G

 Warning of a potential danger, comply with 
instruction manual.

F

 Caution! Dangerous voltage. Danger of 
electrical shock.

 Continuous double or reinforced insulation 
complies with category II DIN EN 61140.

 Equipment  for  working under live voltage

G

 Safety Warnings

G

 Before commencing any testing, short the 
prods together to check the batteries.

F

 The TPT420 is designed to offer a high degree 
of safety. It conforms to standards DIN VDE 
0682-401 and IEC/EN 61243-3. To ensure 
that the unit is operated safely, please read 
the following instructions before using the 
device.

 

Do not use 

the instrument in a humid  

environment. 

 

Before testing a live circuit the voltage 

tester 

should

 be checked by connecting 

to a known voltage source (e.g. a suitable 

120 / 230 V outlet or a suitable proving unit) 

before and after use.

 

Tests for voltage 

should

 be made with both 

poles.

 

Extreme care 

must be taken when testing 

voltages above 50 V AC / 120 V DC. 

 

Ensure the tester is only used within the 

stated measurement ranges, in low voltage 

installations up to 1000 V AC and 1500 V DC 

and in the measuring category it has been 

designed for.

 

When testing, ensure hands are behind the 

barriers on the main body and remote probe.  

Avoid touching the tips. 

 

The instrument must be in good working 

order and be clean and dry, with no broken 

or damage to the  leads, probes or case.

 

Do not use 

the unit with the battery cap 

removed and ensure when changing the 

batteries no probe is connected to any 

circuits.

 

Do not use 

the instrument if one or more 

functions are not working correctly.

 

When testing, the warning triangle “

F

“ 

LED will illuminate, even in the event of the 

Summary of Contents for TPT420

Page 1: ...TPT420 Two pole voltage tester User guide Test Equipment Depot 800 517 8431 5 Commonwealth Ave MA 01801 TestEquipmentDepot com...

Page 2: ...by connecting to a known voltage source e g a suitable 120 230 V outlet or a suitable proving unit before and after use Tests for voltage should be made with both poles Extreme care must be taken whe...

Page 3: ...age is present When there is no indication of voltage present it is still recommended to earth equipment before commencing work A voltage detector of relatively high internal impedance compared to the...

Page 4: ...rotected by an RCD RCBO and Safety Breaker the TPT420 will not trip these devices The phase rotation indication function is simplified which avoids the crossing of the test probes adopted by some 2 po...

Page 5: ...an also be switched on with the torch light button Auto power off The tester is automatically powered off after 5 sec when there is no signal detected by the probes The torch light switches off after...

Page 6: ...age warning LED illuminates 50 VAC 120 VDC and the LCD display is blank then check the batteries This feature must not be used as a means of continuing voltage indications If the low battery indicator...

Page 7: ...ED indicates DC DC During a voltage test L or R LEDs may illuminate For voltages over 1000 V AC and 1500 V DC the LCD will display OL Single pole phase test Hold the tester securely in your hand Conne...

Page 8: ...t good enough Torch light Pressing the torch light button will turn on the light and after approximately 10s it will turn itself off Battery Replacement Remove the probes from any testing point when o...

Page 9: ...r of Electrical and Electronic equipment The registration no is WEE DJ2235XR Users of Megger products in the UK may dispose of them at the end of their useful life by contacting B2B Compliance at www...

Page 10: ...ange indication OL Peak current Is 3 5 mA at 1000 V Measurement Duty 30 s ON operation time 240 s OFF recovery time Internal battery consumption Approx 80 mA Single pole phase test voltage range 100 1...

Page 11: ...a Returns Authorisation number must first be obtained by contacting one of the addresses shown You will be asked to provide key information such as the instrument serial number and fault reported when...

Page 12: ...s user guide is in compliance with Directive 2014 53 EU Other equipment manufactured by Megger Instruments Limited described in this user guide is in compliance with Directives 2014 30 EU and 2014 35...

Page 13: ...M Test Equipment Depot 800 517 8431 5 Commonwealth Ave MA 01801 TestEquipmentDepot com...