SPECIFICATION
WM6
Ranges
insulation
0–200 M
Ω
and
∞
continuity
0–100
Ω
Terminal Voltage
insulation range
<600 V
on Open Circuit (d.c.)
continuity range
800 mV approx.
Terminal Voltage at
insulation range
500 V + 10% –5%
1 M
Ω
Load (d.c.)
Terminal Current
insulation range
1,3 mA approx.
on Short Circuit
continuity range
55 mA approx.
Voltage Stability
insulation range
<±1%
160 r.p.m. to 240 r.p.m.
Ripple Content at 160 r.p.m.
insulation range
500 mV pk-pk approx.
on Open Circuit
Accuracy
1,5 mm (0,060 in) from any marked position on the
scale when measured against standard resistors.
Temperature
operating
–10 °C to +50 °C
storage
–20 °C to +70 °C
Flash Test
2,3 kV a.c. r.m.s.
Voltage Source (d.c.)
Brushless a.c. hand-cranked generator with
rectifier and d.c. to d.c. converter.