MT-10 

Series

Compound Microscope 

INSTRUCTION MANUAL

Use of the S100x oil-immersion objective

The MT-10 series microscopes are equipped with a S100x N.A. 1.25 oil immersion 

objective. Please follow these instructions for using this objective:

• 

Focus the image with the S40x objective.

• 

Turn the revolving nosepiece so the S100x objective almost reaches the 

click-stop.

• 

Put a small drop of immersion oil on the centre of the slide.

• 

Now turn the S100x objective so that you feel the click stop.

• 

The front lens is in contact with the immersion oil. 

• 

Look through the eyepiece (B) and focus the image with the fine 

adjustment knobs (I).

• 

The distance between the lens of the objective and the slide is only 0.14 

mm!

• 

In case there are small bubbles visible turn the S100x objective a couple of 

times left/right so that the front of the objective moves in the oil and the bubbles 

will disappear.

• 

After using the S100x objective turn the fine adjustment knobs (I) up until 

the front lens doesn’t touch the oil any longer.

• 

Always clean the front lens of the S100x objective with a piece of lens paper 

that is moistened with a drop of xylol or alcohol.

• 

Clean the slide after use as well.

Caution

• 

Never put a drop of xylol or alcohol directly on the lens of the objective. It 

 

could enter the objective and dissolve the glue that holds the lenses!

• 

Avoid oil contact with any of the other objectives!

Summary of Contents for MT-10 Series

Page 1: ...MT 10 Series Compound Microscope INSTRUCTION MANUAL...

Page 2: ...roscope INSTRUCTION MANUAL Table of Contents 1 Introduction 2 3 Construction of the microscope 4 5 Functions of the microscope 6 Preparing the microscope for use 7 9 Working with the microscope 10 11...

Page 3: ...microscopes are developed for use at schools and laboratories The maintenance requirement is limited when using the microscope in a decent manner This manual describes the construction of the microsco...

Page 4: ...e mono 360 rotatable B Eyepiece C Stand arm D Revolving nosepiece for 4 objectives E Objectives F Safety device G Object stage with clips H Coaxial coarse and fine adjustment I Lamp housing J Light in...

Page 5: ...MT 10 Series Compound Microscope INSTRUCTION MANUAL A B C D E F G H I K M N...

Page 6: ...ifications and type Achromatic 4x 10x 40x LED illumination Achromatic 4x 10x S40x S100x LED illumination Achromatic 4x 10x S40x S60x LED illumination Achromatic 4x 10x S40x LED illumination Digital he...

Page 7: ...ine adjustment knobs there is a graduation with intervals of 0 002 mm This can be used to measure depths in a specimen Abbe condenser with iris diaphragm Beneath the object stage an Abbe condenser L N...

Page 8: ...ope for Use Remove the packing and put the microscope on a flat table The objectives are pre mounted Put the plug into the mains supply and switch on the microscope Sit comfortably down behind the mi...

Page 9: ...hragm until it is just visible on the edge of the field of view The microscope is properly set for use with the 4x objective For each other magnification this procedure should be repeated to ensure th...

Page 10: ...ge with the fine adjustment knobs I The distance between the lens of the objective and the slide is only 0 14 mm In case there are small bubbles visible turn the S100x objective a couple of times left...

Page 11: ...d of view it resides on the lowest lens of the eyepiece By using the Allen key the eyepiece can be removed from the tube Clean the outside of the lens In case there is still dust visible please check...

Page 12: ...ging the batteries of the MT 30 MT 90 series Caution Always remove the power cable from the main supply Remove the smal lid out of the bottom cover of the microscope Place the batteries and put the li...

Reviews: