Func onal Details
Front panel components
Figure 3. Front panel
1
Screw terminal
2
Spring terminal
Screw terminal
Connect up to four digital I/O lines (DIO0 to DI03) to the detachable screw terminal.
Figure 4. Screw terminal pinout
The digital ground (GND) terminals provide a common ground for the digital bits.
Digital connec ons
The digital I/O lines are electrically isolated from the thermocouple circuits. Each bit is individually configurable
for input (trigger) or output (alarm), and features Schmi trigger inputs and open drain outputs. All DIO
channels are pulled high to 5 V.
Any digital bit can be configured with the web interface as a trigger to start or stop an acquisi on. Any digital
bit can also be configured to trigger an alarm, and be driven high or low when an alarm occurs.
Spring terminal
Connect up to 16 thermocouples (TC0+/TC0– to TC15+/TC15–) and two common reference connec ons (COM)
to the 36-posi on detachable spring terminal. Refer to Figure 5 for signal loca ons.
Cau on! We strongly recommend that you ground yourself using a wrist strap before handling
thermocouple sensors.
RedLab WebDAQ 316 User's Guide