3.5.2 Operating the display screen

3

.

5 Function description

Ø

Oxygen flux display: in L/min;

Ø

Oxygen concentration display: in %(V/V);

Ø

Oxygen generation icon: when the icon is on, display the oxygen generation status;

Ø

Timing display: display the current use time and set the single use time;

Ø

Accumulated timing function: display the accumulated running time of the oxygen generator;

Ø

Alarm  function:  including  power-off,  oxygen  concentration  detection  function,  corresponding 
indicator flashing or warning voice sounding for reminder in case of a fault;

Ø

Mute icon: when the icon is on, the key tone will be enabled; otherwise, the key tone 
will be disabled;

Ø

Atomization function: compressed air atomization (optional);

Ø

Remote control function (optional).

Product Model

ZY

-

F1WA2

ZY

-

F2WA2

Rated flux

1L/min

2L/min

Input power

130VA

180VA

Net weight

6.5kg

8.0kg

Noise

=45db

=50db

Oxygen concentration

≥90%(about 30min later after startup)

Flow adjustment range

1

6L/min

Atomization rate

=

0.2mL/min

Exterior dimensions

Ф

278x39

8

mm

Outlet pressure

30

60KPa

When the nominal outlet pressure is 7kpa, the flux level will be 1~6L/min;
When the maximum flux is recommended, by applying 7 kPa of back pressure, the flux  will change 
by 0.5 L / min;

OXYGEN FLOW

CONCENTRATION

L/min

%

TIME

Min

H

TOTAL TIME

OXYGEN

FAULT

MUTE

MUTE

TIMING

TIMING

ON/OFF

FLOW

Oxygen Flow

Oxygen 
generating icon

Timing display

Sound icon

Accumulated 
timing

Sound button

Oxygen 

concentration

Infrared receiver

Fault icon

On-Off button

Flow button

Timing -

3.6 Technical specifications

3.7 Working environment

Ø

Ambient temperature range: 5°C~40°C;

Ø

 Relative humidity range: ≤80%;

Ø

 Atmospheric pressure range: 86kPa -106kPa;

3

Ø

 Impurities in feed air: ≤3mg/m  

3.8 Storage and transportation environment 

Ø

 Ambient temperature range: -20°C~+55°C;

Ø

 Relative humidity range: ≤95%;

Ø

 Atmospheric pressure range: 70kPa -106kPa;

Ø

 When  the  storage  and  transportation  temperature  is  lower  than  5  °C,  the  equipment 

shall be kept in the normal working temperature environment for more than 4h before 
use.

Ø

The oxygen generator shall be stored in a well-ventilated room without strong sunlight 

or  corrosive  gas.  During  transportation,  it  shall  be  prevented  from  strenuous  vibration 
and lying upside down or horizontally.

Summary of Contents for ZY-F1WA2

Page 1: ...ompany Limited Address No 1862 Zishi Road Hefei City Anhui P R China Production Address No 1862 Zishi Road Economic and Technological Development Zone Hefei City Post Code 230601 Email zkmeiling zkmei...

Page 2: ...am After reading this manual please keep it properly for your reference at any time The warranty card is attached and shall not be lost See the warranty card for the date of production The figures in...

Page 3: ...y in terms of oxygen flux and oxygen absorption time please use it under the guidance of a doctor To prevent power failure or possible failure of the oxygen generator users in urgent need of oxygen an...

Page 4: ...cal institutions and families use it to produce oxygen for oxygen supply to patients with hypoxia 3 2 Effects of oxygen By supplying oxygen to patients it can cooperate with the treatment of cardiovas...

Page 5: ...ressure 30 60KPa When the nominal outlet pressure is 7kpa the flux level will be 1 6L min When the maximum flux is recommended by applying 7 kPa of back pressure the flux will change by 0 5 L min OXYG...

Page 6: ...flux When the oxygen generator is working the sound of puff will be emitted every few seconds which is the normal reversing exhaust sound At the beginning of startup the oxygen concentration will incr...

Page 7: ...ease timing every 30min with a 0 240min cycle When the timing is in working state the time will be displayed by countdown and refreshed once every minute When the remaining time is within 3min and the...

Page 8: ...the products that have obtained the registration certificate of medical devices Note the accessories recommended by the manufacturer shall be used otherwise the product performance may be affected 7 T...

Page 9: ...he oxygen concentration will not decrease and the efficiency is less than 90 when the sea level is from 1828m to 4000m When the nominal outlet pressure is 0 the change of oxygen concentration and oxyg...

Page 10: ...anual including warranty card Certificate of Conformity Number 1 set 1 pc 1 unit 1 unit 1 pc 1 pc 2 pcs 1 1 The circuit and appearance of the equipment are subject to change for improvement without pr...

Page 11: ...y service and its compressor can enjoy three years free warranty service with the purchase invoice 2 We will not provide free warranty service for faults caused by the following personal reasons of us...

Reviews: