Hard Reset
Press and hold the power button for 30 seconds.
Audio Signals
Two Beeps
– Invalid rotation during measurement.
Two Long Beeps
– Measurement timeout or bump warning.
Single Beep
– Powering on or returning from an invalid rotation.
Single Low Pitch Beep
– Button press.
Single Long Beep
– Powering off.
Error Modes
Invalid Rotation
– During a measurement, it is possible
for the device to be oriented in an invalid direction (see
Measurement Guidelines). When this happens, the display
shows a spirit level where the bubble indicates how the
device should be redirected to get back to a valid direction.
The measurement is resumed when the bubble gets close
enough to the center circle.
Bump Warning
– If the device is bumped or moved too
fast during the measurement the acceleration error screen
is shown and the measurement is aborted. Press the main
button to restart the measurement.
Measurement Timeout
– If a measurement takes more
than one minute it is aborted and the timeout screen is
shown. Press the main button to restart the measurement.
59º
121º
2º 178º
42º 138º
9º 171º
26º 154º
59º
2º 178º
42º 138º
9º 171º
26º 154º
2º
178º
42º 138º
9º 171º
26º 154º
Delete measurement
Previous measurements can be deleted in the history view. Pressing
the delete button highlights the latest measurement, pressing the delete
button again confirms deletion. Pressing any other button cancels the
deletion.
EN - EasyAngle - Instructions for Use
History View
Delete Pressed
Delete Confirmed
Indicators
There are two indicators on the display, the battery indi-
cator and the workflow indicator. The battery indicator
has four levels, please charge the battery when the last
level is displayed. The workflow indicator shows which
step in the workflow is currently active, see the figures in
Basic Operation
.
33º
147º
7º 173º
44º 136º
Battery
Workflow
3