54
Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
6.4.5.2 Remote Communication Settings
The remainder of the Parameters in the Remote Application present the choice of a numeric entry, a
numeric address entry, or a radio button selection. The parameters that require a numeric entry will
present a numeric keypad with min and max limits for the variable. The parameters that require a numeric
address (for example an IP address) will also present a keypad. Address entries should conform to the
format of the parameter selected. The three screens below show examples of each type.
Numeric address entry
Figure 6.4.5.2 - Remote Communication Settings
Numeric entry
Radio button selection
i
Setting the Ethernet DHCP to
yes will have a short delay while
the DHCP server is contacted.
If a DHCP server is not found,
an error will be indicated. If DHCP is
enabled, the IP address, Netmask and
Gateway are greyed out and locked, these
are controlled by the DHCP server.