Operating Instructions - CPC6050
91
Modular Pressure Controller
CPC6050
:INDEX?
Returns the index number.
index unit
0 bar
1 mbar
2 Pa
3 psi
4 atm
5 kp/cm2
6 lbf/ft2
7 kPa
8 cmH2O(4°C)
9 inH2O(4°C)
10 inH2O(60°F)
11 ftH2O(4°C)
12 µmHg(0°C)
13 mmHg(0°C)
14 cmHg(4°C)
15 inHg(0°C)
16 inHg(60°F)
17 - -
18 user
19 user
20 user
index unit
21 OSI
22 TSI
23 TSF
24 ftH2O(20°C)
25 ftH2O(60°F)
26 inSW
27 ftSW
28 hPa
29 Mpa
30 dyn/cm2
31 mHg 0°C
32 Torr
33 mTorr
34 uHg 0°C
35 mmH2O 4°C
36 mH2O 4°C
37 mmH2O 20°C
38 cmH2O 20°C
39 mH2O 20°C
40 mSW
41 g/cm2
42 Kg/m2
OUTPut
:STATe ON | OFF | 1 | 0
ON or 1 = Control OFF or 0 = Measure
:STATe?
Returns 0 for measure 1 for control
:MODE MEASure | CONTrol | VENT
Sets the mode indicated
:MODE?
Returns the mode string
:STABle?
Returns 1 if stable 0 if not
:AUTOvent ON | OFF | 1 | 0
Autovent is always true on CPC6050
:AUTOvent?
Returns the state of the Vent mode
:AUTORange ON | OFF | 1 | 0
Turns autorange on or off
:AUTORange?
Returns “ON” or “OFF”
:CHANnel A, B, D
Sets the active channel
:CHANnel?
Returns the active channel
[SOURce]
:PRESsure
[:LEVel]
[:IMMediate]
[:AMPLitude] <n>
Sets the setpoint