Operating Instructions - CPC6050
19
Modular Pressure Controller
CPC6050
4.2 Base Instrument
Instrument
Instrument version
Standard: desktop case
Option: 19" rack-mounting kit
Dimensions
See technical drawings
Weight
approx. 22.7 kg (50 lbs) incl. all internal options
Warm-up time
approx. 15 min
Display
Screen
8.9" color LCD with resistive touchscreen
Resolution
4 ... 6 digits depending on range and units
Connections
Pressure connections
up to 8 ports with 7/16”- 20 F SAE, up to 2 ports with 1/8” F NPT and 1 port with 10-32 UNF female
Filter elements
The instrument has a 40-micron filters on all pressure ports.
Pressure port adapters
Standard: without
Option: 6 mm tube fitting, 1/4" tube fitting, 1/4" female NPT fittings, 1/8" female NPT fittings or 1/8" female
BSP fittings
Barometer port adapters
Standard: Barb Fitting
Option: 6mm tube fitting, 1/4“ tube fitting
Permissible pressure media
Dry, clean air or nitrogen (ISO 8573-1:2010 class 5.5.4 or better)
Wetted parts
Aluminum, brass, 316 and 316L stainless steel, Buna N, FKM/FPM, PCTFE, PEEK, PTFE, PPS,
glass-filled epoxy, RTV, ceramic, silicone, silicone grease, Urethane
Overpressure protection
Safety relief valve fixed to reference pressure transducer and adjusted to customized measuring range
Permissible pressure
Supply port
110% FS or 0.69 bar (10 psi), whichever is greater
Measure/Control port
max. 105 % FS
Voltage supply
Power supply
AC 100 ... 120 V, 50/60 Hz; AC 220 ... 240 V, 50/60 Hz
Power consumption
max. 210 VA
Permissible ambient conditions
Storage temperature
-20 ... 70 °C (-4 ... 158 °F)
Humidity
5 … 95% r. h. (relative humidity non-condensing)
Compensated temperature range
15 ... 45 °C (59 ... 113 °F)
Mounting position
horizontal
Control parameter
SVR module
8)
LPPump module
Control stability
< 0.003% FS of the active range
(typical 0.001% FS
9)
)
< 0.003% FS of the active range
(typical 0.001% FS
9)
)
Control mode
precision, high speed and custom
external supply on / off
Control time
15 s
10)
25 s
10)
Control range
0 ... 100% FS
0 ... 100 % FS
Minimum control pressure
0.0017 bar (0.025 psi) over exhaust pressure or
0.05% FS whichever is greater
0.0034 bar (0.05 psi) over exhaust pressure or
0.05% FS whichever is greater
Overshoots
< 1% FS in high speed control mode (typical
<0.05% FS in precision control mode)
< 1% FS in high speed control mode (<0.1% FS
in pump only mode)
Test volume
50 ... 1,000 ccm
50 ... 300 ccm
Communication
Interface
Standard: Ethernet, IEEE-488, USB, RS-232.
Command sets
Mensor, WIKA SCPI, others optional
Response time
approx. 100 ms
Internal program
up to 24 sequences with up to 99 steps each
8) Represents LPSVR, MPSVR, HPSVR and EPSVR
9) Typical stability achieved 10 seconds after the stable indication, when controlling on pressure above atm
10) Regarding a 10 % FS pressure increase above atm. in a 50 ml test volume, in high speed mode (SVR) or external supply on (LPPump)