6.4.3.8 Vent Rate
The Vent Rate button (SVR Module only) in the Control Settings App lets the user to determine the rate
at which pressure will vent in vent mode. By default the rate of the vent is set similar to the control rate.
Figure 6.4.3.8 displays the vent rate setting for channel A.
Figure 6.4.3.8 – Vent Rate Setting
6.4.3.9 Rate Stability Parameters
Rate parameters for the control rate (SVR Module only) can be found in the Control Settings App and can
be configured using the Rate Stable Window and Rate Stable Delay buttons. The Rate Stable Window
button allows the user to enter a value as a percentage of the active range of transducer. This value rep-
resents the pressure window within which the control rate value would be considered stable by the user.
The Rate Stable Delay button lets the user add a desired delay until the control rate is considered stable
while being in the rate stable window.
Figure - 6.4.3.9 Rate Stability Parameters