48
Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
6.4.3.10
Detection Flags
The CPC6050 is equipped with three detection flags (SVR Module only) that can be enabled or disabled
by the user as needed. These detection flags appear in the Control Settings App. The primary purpose
of these detection flags is to protect the instrument and to ensure desired operation. Each of these flags
can be turned “On” or “Off” as per user’s need. Figure 6.4.3.10 shows the detection flag settings.
• Supply Detection Enable: If turned “On”, this flag allows the user to check for sufficient supply pres-
sure at the Supply Port of the instrument. In case the supply pressure is less than 10% of the control
setpoint, an error is reported which can be seen in the Troubleshoot App by clicking the error symbol
[ ]. This flag is turned “Off” by default.
• Burst Detection Enable: If turned “On”, this flag allows the user to protect the instrument against sud-
den bursts in the pressure at the Measure/ Control Port. In case a burst is detected, an error is report-
ed which can be seen in the Troubleshoot App. This flag is turned “On” by default.
• Measure Regulation Enable: If turned “On”, this flag allows pressure to be controlled while the instru-
ment is in “Measure” mode to prevent pressure leaks in the system over time. This will turn on the in-
ternal pressure regulator periodically to control pressure in a certain range of the measured pressure
value. This flag is turned “Off” by default.
Detection
Flags
Figure 6.4.3.10 - Detection Flags