Operating Instructions - CPC6050
55
Modular Pressure Controller
CPC6050
6.4.6 Step Settings Application
The Step Settings Application allows the user to enter the minimum and maximum pressure
values of the device under test (DUT) and the number of preset steps required across the span.
From this information the Step Settings App will automatically calculate setpoint values in pres-
sure units and values as a percentage of the full span of the DUT (figure 6.4.6). To enter the
minimum and maximum range of the DUT select the DUT Maximum Range and DUT Minimum
Range buttons (figure 6.4.6-A) and enter the values. These ranges should be within the maxi-
mum range of the primary transducer in the selected channel. CPC6050 automatically switch
between Pressure units and Percent of DUT Range values by pressing the units button or the
percent button (Figure 6.4.6 - B). The user can be directed to this application from the setpoint
entry method in the Home App (section 6.4.1.2). The Overrange button allows an entry above
100% in order test for hysteresis on a dial gauge.
Figure 6.4.6 – Step Settings Application
Figure 6.4.6-A – DUT Range Selection Figure 6.4.6-B - Units or Percent Selection