High-End Pressure Controller
CPC8000
Operating Instructions - CPC8000
13
3. Product Description
3.1 General Description
The CPC8000 High-End Pressure Controller is our highest performance pressure controller and features
the following:
• 19” rack mount compatibility
• Designed for serviceability and reliability
• Widescreen display with glass capacitive touchscreen
• Front panel door gives access to instrument interior
• Up to three removable sensors 0.01% IS-
50
(IntelliScale-
50
)
• Pressure ranges from 10.36 pressure regulator, up to 10:1 turndown
• Adaptive control algorithm with rate control capability
• Ethernet, RS-232, USB, and IEEE-488 communications
• Remote compatibility with PCS 400, CPC 6000, CPC 3000, old CPC8000 (SCPI)
• Emulation of competitive instrument command sets
• Local program capability
• Multiple language support
• On board diagnostic logging
• Barometric reference option
• 0.008% IS-
33
(IntelliScale-
33
) sensor option
• Complies with latest CE and 61010 regulations
The CPC8000 is designed to automate the testing and calibration of all types of pressure devices and
instruments. Up to three removable / interchangeable pressure transducers are available in full scale
(FS) ranges from 0.36 to 6,000 psig, 7.5 to 6015 psia. Each transducer module is confgured with its own
calibration parameters on board and have 0.01% IS-
50
or optional 0.008% IS-
33
uncertainty.
The three transducers, in combination with the internal control valve regulator, provide a dynamic output.
The operator can choose to control pressure either using a single, selected transducer or auto-range
control across all three sensors. The ratio between the highest FS range and the lowest FS range within
the CPC8000 cannot exceed 10:1. The three ranges can be chosen to optimise uncertainty levels across
the full pressure span of the instrument.
In addition to the capacity for three active ranges, a fourth, barometric transducer is available as an
option. With this option installed a CPC8000, with absolute or gauge pressure transducers, can emulate
pressure of the opposite type.
3.1.1 IntelliScale Transducer Choices
The CPC8000 can be supplied with standard 0.01% IS-
50
Transducers or optional 0.008% IS-
33
trans-
ducers to give a percent of reading uncertainty down to 50% or 33% of full scale respectively. Refer to
, for a full description of the uncertainty specifcation and
.
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...