4. Specifcations
Accuracy specifcations presented herein are obtained by comparison with primary standards traceable to
the National Institute of Standards and Technology NIST). A calibration data report has been provided with
your new CPC8000 instrument, traceable to NIST. The calibration program at Mensor is accredited to both
ISO/IEC 17025:2005 and Z540-1-1994 by A2LA. Mensor, LP is registered to ISO 9001:2008.
Mensor reserves the right to change these specifcations without notice.
4.1 Measure Specifcations
Total Uncertainty
(1)
0.01% FS
(3)
, 0.01% IS-50
(2)
, 0.01% Span, 0.03% FS, 0.03% Span
-or-
Optional 0.008% IS-
33
(4)
Optional 0.008% IS-50
(5)
Gauge and absolute emulation modes add or RSS the additional uncertainty of the barometric
transducer (0.01% of reading).
Pressure ranges
0 ... 7.5 psia to 0 ... 6015 psia
0 ... 0.36 psig to 0 ... 6000 psig
-15 ... 1015 psig to -15 ... 6015 psig (0.01% IS-50 Uncertainty
(2)
)
-5 ... 5 psig to -15 ... 145 psig bi-directional (0.01% Full Span
(3)
)
See Table 4.6.1 for available Transducer Type, Range, Mode & Uncertainty)
Maximum ratio between highest and lowest range is 10:1
Measurement units
psi, inHg@0C, cmHg@0C, mmHg@0C, inH2O@4C, 20C, bar, mbar, cmH2O@4C and 20C, hPa, kPa,
Pa, MPa, kg/cm sq, and 2 user-defned units. All above in rate (per second, per minute and per hour).
Resolution
Operator selectable from 4 to 6 digits
Calibration adjustments
Zero adder and Span multiplier or up to 11 point linearization for each sensor
Calibration interval
365 days for IS-33 or IS-50 and 180 days for standard transducers, with periodic re-zeroing.
Calibration data storage
The calibration data is stored on the sensor.
Measurement flters
Of, low, normal (default), high
Defnitions: Max = the maximum value of a range, also known as the full scale value. Example, for a range of –15 ... 145, Max = 145
Min = the minimum value of a range. Example, for a range of –15 ... 145, Min = –15
Full Span = Max – Min. Example, for 0 ... 145 the Full Span is 145, for –15 ... 145 the Full Span is 160
Reading = The value of the sensor output
(1) Total Uncertainty (k=2) includes hysteresis, linearity, repeatability, reference standard, drift and temperature efects over the calibrated range for the calibration interval speci-
fed, with periodic re-zeroing.
(2) 0.01% Intelliscale-50 (0.01% IS-50): Uncertainty from Min to 50% of Max = (0.010% x 50% x Max) or 0.010% of reading from 50% to 100% of Max for 365 days.
(3) 0.01% Full Span (0.01% FS) = Uncertainty of 0.01% of the Full Span for 365 days.
(4) 0.008% Intelliscale-33 (0.008% IS-33): Uncertainty from 0 to 33% of Max= (0.008% x 33% x Max) and 0.008% of Reading from 33% to 100% of Max for 365 days.
(5) 0.008% Intelliscale-50 (0.008% IS-50): Uncertainty from 0 to 50% of Max = (0.008% x 50% x Max) and 0.008% of Reading from 50% to 100% of Max for 365 days
4.2 Control Specifcations
External pressure
requirements
Source Requirements – 10% over range of highest pressure transducer
Exhaust Requirements – Vacuum source required for any control below atmosphere
Control stability
0.002% FS of highest range sensor at the set point
Pressure range ratio
Maximum ratio between highest and lowest range is 10:1
Minimum control pressure
0.05% FS of highest range sensor or .025 psi over exhaust pressure, whichever is greater
Pressure control rates
0.001% of range/sec to 10% of range/sec
Control time
60 seconds to stable fag for a 10% pressure change into 150cc volume. Larger volumes can
lengthen this time. Controlling to pressures less than 0.5 psia will lengthen this time.
Supply consumption
<2.5 scfh in steady-state control.
Overshoot
<0.1%
External volume
50 to 1,000 milliliters. Requires one complete pressure cycle after changing the external volume,
or after a power cycle. See page 27, “Turning on the CPC8000”.
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...