High-End Pressure Controller
CPC8000
Operating Instructions - CPC8000
27
5.4 Turning on the CPC8000
After the pressure connections are secure in accordance with Section 5.5 below, apply power to the power
connector on the rear of the instrument and switch the power switch ON. The instrument will go through an
initialization process and system check. As soon as the system check is completed the system will default
to an operating screen similar to the screen shown in
. Allow the instrument at least
30 minutes of warm up time before performing critical pressure measurements.
Earth Ground! Any power adaptors or surge protection devices that negate the
protective earth ground should not be used. The power cord must be accessible
and contain a protective earth ground.
Ventilation! Do not block airfow to ventilating fans located on rear of instrument.
The instrument can be set up on a table top or it can be rack mounted. Rack mount adapters are optional.
After the controller has warmed up, cycle the pressure from the minimum to the maximum then back to
the minimum pressure of the primary sensor with the desired volume connected to the Measure/Control
port. This will allow the controller to properly compensate for the diference in external volume between
what was used at the factory versus what is currently in use. This pressure cycle should occur after each
power cycle, or when the external volume size has been changed.
5.5 Pressure Connections
Warning! The pressure connections must be installed according to the following
instructions, observing the relevant regulations. The installation is to be performed
by trained, authorized personnel, knowledgeable in the safety regulations for
working on pneumatic/hydraulic systems.
Up to 6 pressure connections are on the rear panel. Pressure connections that are not
assigned are plugged.
SUPPLY
EXHAUST
MEASURE\
CONTROL
VENT
REF
BAROMETRIC
REFERENCE
Figure 5.5 - Pressure Connections
High-End Pressure Controller - CPC8000
Power supply: 100 ... 120 VAC / 200 ... 240 VAC, 50 ... 60 Hz
Energy consumption: max. 140 VA
Pressure medium: clean dry air or nitrogen
Operating temp. : 15 ... 45°C
Serial number: 910021
Basic instrument: LP-NVR (-14.5 … 90 psi / 0 … 105 psi abs.)
Date of manufacture: 12/2012
Mensor - 201 Barnes Drive - San Marcos, TX 78666
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...