High-End Pressure Controller
CPC8000
36
Operating Instructions - CPC8000
Figure 6.2.1 - [Autorange] / [Range Hold]
6.2.2 Pressure Reading
The large, green number displayed in fgure 6.2.1 is the pressure reading at the sensor connected to the
Measure/Control port. It is green because the sensed pressure is within a user-defned stable window
for a user-defned period of time. When the pressure is outside of the stable window, the number will be
white.
for a detailed explanation and setup of the stable window.
The pressure reading will turn red if an external pressure is applied at the Measure/Control
port that is 7% greater than the upper limit of the active transducer. Each transducer is sup-
plied with a pressure release valve set for 10% over the calibrated pressure. Inherent internal
safeguards are set to prevent overpressure.
6.2.3 [Setpoint] Button
The [Setpoint] button will activate a sidebar data entry method to accept a new setpoint value. The
setpoint value is displayed in the window just below the [Setpoint] button on the main screen,
. The setpoint value can also be adjusted incrementally in the sidebar by use of one of four setpoint
entry methods:
■
The numeric key pad
■
A numeric step value assigned to the up [▲] and down [▼] arrow keys
■
A digital step assigned to the up [▲] and down [▼] arrow keys
■
A program
Procedures for making changes to the setpoint using these methods are covered in
.
6.2.4 Setpoint Value
The setpoint value window is located immediately under the [Setpoint] button. It displays the current
pressure setpoint (target value for the output pressure provided by the internal regulator when in Control
mode). To edit the setpoint value, frst touch the [Setpoint] button. This will activate the system for setpoint
changes. The sidebar will display the default number key pad or a pre-chosen setpoint entry method. To
change the setpoint value use one of these methods described in
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...