6.2.10 Limits
Limits are shown in three places in the main screen. The “Range Limits”, “Control Limits” and “Entry
Limits” shown in fgure 6.2.10 below can show two diferent readings depending on settings. The Range
Limits shows the limits of the range selected in Range Hold mode or the limits of the active transducer
in Autorange mode. The “Control Limits” are the limits entered into the Setup [
] / [Control] [Maxi-
mum Limits]/[Minimum Limits] screen (
). The “Control Limits” limit the value that can be
entered as a setpoint. The “Entry limits” are equal to the lessor of the Control Limits or the Range Limits
(when in Range Hold mode).
Control
Limits
Entry
Limits
Range
Limits
Figure 6.2.10 - Limits
6.2.11 Bar Graph
The bar graph is always active and reports in real time the pressure at the Measure/Control port. This
provides a visual comparison between the actual pressure and the user selected control limits.
Bar Graph
Figure 6.2.11 - Bar Graph