High-End Pressure Controller
CPC8000
48
Operating Instructions - CPC8000
6.2.14.2 Control Mode
In Control mode, the instrument provides a precise pressure output at the Test port. The indication of
the current pressure value will turn green when the setpoint has been reached and the stable window
settings have been satisfed.
Control Mode
Figure 6.2.14.2a - Control Mode
The Control mode is activated by pressing the [Control] button in the main menu. The Control mode can
be accessed from the Measure mode or the Vent Mode by pressing the [Control] button. In the Control
mode the CPC8000 acts as a precision pressure controller and provides a stable pressure output at the
Measure/Control port.
In order to ensure that the controller is correctly confgured for the task it is to perform, the following
measures must be taken and the respective parameters must be set in the Setup [
] menu.
• In order to control pressures close to or below atmospheric pressure, a vacuum pump should be
connected to the Exhaust port.
• The control speed can be set in the Setup [
] / [Control] / [Rate setpoint] screen. The Control rate
can be set between 0.001% of range/sec to 10% of range/sec.
• Control limits can be set in the Setup [
] / [Control] / [...limits] screen.
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...