6.2.14.3 Vent Mode
Vent mode vents the pneumatic system and shuts of the supply. The Vent mode can be activated from
the Measure or Control mode by pressing the [Vent] button. Internal system pressure will be vented
through the Vent port.
Vent Mode
Figure 6.2.14.3a - Vent Mode
WARNING! Venting will cause a sudden loss of pressure in the system and the
plumbing connected to the measure control port. Care must be taken that the
device under test is not damaged during venting.
WARNING! In some cases venting will exhaust high velocity air from the vent port.
Personnel and sensitive equipment should not be in close proximity to the vent port
during venting.
WARNING! In some cases venting high pressure will cause a loud sound. Personnel
should wear a hearing protection device when working with high pressures.