High-End Pressure Controller
CPC8000
56
Operating Instructions - CPC8000
7.1.5 Load (Confguration)
Each confguration can store a complete set of parameters and settings. Confgurations can be recalled
(loaded) as needed. Confgurations contain settings for the following set of parameters:
Lower limit
Upper limit
Setpoint
Step
Rate setpoint
Rate step
Stable window
Stable delay
Rate stable
window
Rate stable delay Vent rate
Emulation mode
(gauge/abs)
Filter
Resolution
Units
Rate units
Secondary units Brightness
Volume
Table 7.1.5 - Confguration Parameters
Explanation on how to set up each confguration can be found below in Section 7.1.6
The [Load] button within the [General] tab of the Setup [
] menu, shown in fgure 7.1.5, provides a
place to choose a saved confguration. When the [Load] button is chosen a choice of several confgura-
tions are displayed on the sidebar. When one confguration is chosen its saved parameters and settings
will be active in the instrument.
Figure 7.1.5 - Load (confguration)