High-End Pressure Controller
CPC8000
92
Operating Instructions - CPC8000
Step-
Jogs the setpoint down one step
Step+
Jogs the setpoint up one step
Step?
<sp>+n.nn<cr><lf>
Returns the control step for the instrument
Step%
Value in % of current range
Sets the control step in % of current range
Steppct
Value in % of current range
Sets the control step in % of current range
Steppct?
<sp>+n.nn<cr><lf>
Returns the current step in % of current range
Tare
<ON,OFF>
Tares the current reading to zero. The same
tare value is applied to all measuring sensors.
Tare?
<sp>+n.nn<cr><lf>
Returns the tare value applied to all sensors
in current units.
Termchar
<CR,LF,CRLF,EOI,DEFAULT>
Changes the termination character for all
command sets. If not using GPIB, EOI will
add CRLF. DEFAULT restores default termi-
nation. For Mensor it is CRLF, and for WIKA
SCPI it is LF.
Units
units code or text (see table
11.4)
Sets the instrument engineering units.
Units?
<sp>CCCC<cr><lf>
Returns the instrument units in a text string
Unlock
None
Releases Acquire locks
See Acquire? And Release?
UpperLimit
Value inside primary xducer-
Range in current units
Sets the upper control limit for the active
transducer.
UpperLimit?
<sp>+n.nn<cr><lf>
Returns the upper control limit for the active
transducer.
Vent
None
Instrument placed in Vent Mode
Vent?
<sp>(YES or NO)<cr><lf>
Returns yes if instrument is in Vent, no if
otherwise
Vent_Limit
Value in current units
This is the point at which controlled vent
stops and the vent solenoid is opened.
Vent_Limit?
<sp>+n.nn<cr><lf>
Returns the vent limit in current units.
Vent_rate
Value in current units
This is the rate the controlled vent uses.
Vent_rate?
<sp>+n.nn<cr><lf>
Returns the vent rate in current units.
Window
Value in current units
Sets the exponential flter window for the
active sensor
Window?
<sp>+n.nn<cr><lf>
Returns the exponential flter window for the
active sensor
Zero
desired pressure or ?
Sets zero to set pressure or for ?, clears
previous value. CALDISABLE must be OFF/
NO.
See Section 7.5.2.5.5, Remote Zero
Zero?
<sp>+n.nn<cr><lf>
Returns zero ofset for active transducer
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...