High-End Pressure Controller
CPC8000
Operating Instructions - CPC8000
97
8.5 Command Emulation
8.5.1 PCS 400 Emulated Commands
Table 8.5 - PCS 400 Emulated Commands
Command
Data
Response/Function
_pcs4 autorange <value>
0 or 1
1 turns autorange on, 0 = of
_pcs4 autorange?
Returns 1 if in autorange, 0 if in range hold
_pcs4 cal a/d
Not used, kept for backwards compatibility
_pcs4 cal atm
Performs pcs400 1 pt cal
_pcs4 cal span <value>
Sets the span of the active transducer to <value>
_pcs4 cal zero <value>
Sets the zero of the active transducer to <value>
_pcs4 cal_disable_of
Enables zero or span calibrations if previously disabled.
_pcs4 cal_disable_on
Prevents zero or span calibrations
_pcs4 ctrl
<value><sp><unitno>
Sets control value – will take efect immediately if instrument
is in control mode.
_pcs4 ctrl?
Returns the current control point in current engineering units.
_pcs4 ctrlmax<value>
Sets maximum control value.
_pcs4 ctrlmax?
Returns current maximum control pressure.
_pcs4 ctrlmin<value>
Sets minimum control value.
_pcs4 ctrlmin?
Returns current minimum control pressure.
_pcs4 emul?
Returns ptype emulation mode
_pcs4 default
Sets default values into instrument.
_pcs4 err?
Returns the error number and description.
_pcs4 exhaustp?
Returns exhaust pressure
_psc4 fltersetting
Sets the flter %
_pcs4 fltersetting?
Returns the flter %
_pcs4 flterwindow
Sets the flter window
_pcs4 flterwindow?
Returns the flter window
_pcs4 func ctrl
<value><unitno>
Instrument placed in control mode at <value> pressure in
<unitno> units.
_pcs4 func emul
Toggles ptype emulation mode
_pcs4 func F1
Toggles ptype emulation mode
_pcs4 func meas
Instrument placed in measure mode.
_pcs4 func stby <unitno>
Instrument placed in standby mode in <unitno> units
_pcs4 func vent <unitno>
Instrument placed in vent mode in <unitno> units
_pcs4 id?
Returns instrument ID.
_pcs4 lang PCS2
Sets command set to PCS 200
_pcs4 list?
Returns range list
_pcs4 opt?
Returns option list (old pcs400 format)
_pcs4 option?
Returns option list.
_pcs4 outform<digit>
Sets output format.
_pcs4 outform?
Returns the current output format.
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...