5. Installation
WARNING: READ THESE INSTRUCTIONS BEFORE INSTALLATION!
The installation location must meet the following conditions (see also
):
■
Ambient temperature: Compensated temperature range 15° to 45° C
■
Humidity: 0-95% relative humidity, non-condensing
■
Flat, horizontal location; secure fixed working surface (desk top model) or proper installation in a
sturdy 19” mounting rack or cabinet
■
Sufficient air circulation must be provided at the rear of the instrument to avoid accumulation of heat
Pressure Supply Requirements:
■
Stable supply pressure 10% higher than the full scale of the controller
■
Permissible media: dry, clean air or nitrogen
■
Vacuum: minimum 50 liters per minute (if required)
5.1 Initial Installation
The initial installation of the CPC8000 includes the following steps: Unpack the system, place it in a
suitable work place, connect it, switch it on and confgure.
A Quick Start Guide has been placed inside the shipping box containing your instrument. This guide
provides a quick reference to the pneumatic and electrical connections and an introduction to the opera-
tor interface. It is intended for experienced operators. All safety precautions within this manual should be
understood and followed.
5.2 Unpacking the system
Your new instrument was subjected to many hours of functional testing before it left the factory. In addition
to testing, the unit was inspected for appearance prior to being packaged for shipment. Upon removal
from its carton please examine the instrument for shipping damage. Report any apparent damage to the
carrier immediately.
Apart from any additional components ordered, a shipment consists of:
■
CPC8000 High-End Pressure Controller
■
Power Cord
■
Operating Manual
■
Quick Start Guide
■
Any options ordered
■
Two manifold seal plates (either attached internally or sent externally with the CPC8000)
■
An envelope containing the Calibration Certificate
■
A 3/16 hex screwdriver
■
A microfiber cloth to clean the glass front panel
■
Muffler, if pressure ranges ≥ 1000 psi (70 bar)
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...