7.5.2.5 Calibration Adjustment Procedure
There are three methods that can be used to calibrate all of the transducers in the CPC8000. A simple
[1-Point Calibration] can be used to adjust a single point (usually the zero point). A [2-Point Calibration]
extends this capability to adjust two points (usually points close to the Zero and Span). The [Linearize]
calibration provides a way to calibrate and linearize the pressure curve using from 1 to 10 points over
the range of each transducer. The Remote Zero Calibration is similar to the 1-Point Calibration but is not
permanent (does not persist through power cycles).
Procedures for use and interaction with all the calibration screens can be found in the following sections
of this manual:
• 1 Point Calibration —
• 2 Point Calibration —
• Head Correction —
• Remote Zero Calibration —
When calibrating the CPC8000, adjust for the diference in level between the
pressure standard and the instrument (head) must be considered. See
to view and adjust the head correction.