High-End Pressure Controller
CPC8000
Operating Instructions - CPC8000
93
8.3 SCPI WIKA Command Set
Table 8.3 - SCPI Commands
Command or Query
Response/Function
MEASure
[:PRESsure] [R]?
Returns the pressure from range R.
:TEMPerature[R]?
Returns the temperature from range R.
:RATE[R]?
Returns the pressure rate/sec from range R.
:BAROmetric?
Returns the barometric pressure.
CALibration
[:PRESsure][R]
:MODE?
Returns 1=calibrated or 0=not calibrated.
:DATE?
Returns date of cal “MM/DD/YY”
:DATE <i,i,i>
Sets date of cal YYYY,MM,DD
:ZERO?
Returns zero ofset.
:ZERO <n>
Sets the zero ofset.
(See Section 7.5.2.5.5, Remote Zero Calibration).
:ZERO:RUN
Start autozero sequence.
(See Section 7.5.2.5.5, Remote Zero Calibration).
:ZERO:STOP
Stop autozero sequence
:ZERO:INITiate?
Returns zero status
:ZERO:INITiate
Ignored
SENSe
[:PRESsure] [R]
:NAME?
Returns sensor name string.
:MODE?
Returns pressure type “ABSOLUTE” or “GAUGE”
:MODE ABS/GAUGE
Sets the pressure type to absolute or gauge.
:ABS?
Returns native sensor type “1” for absolute, ”0” for gauge.
:RESolution?
Returns resolution (foat).
:RANGe
[:UPPer]?
Returns the maximum range.
:LOWer?
Returns the minimum range.
:Unit
[:NAME]?
Returns ASCII units (mixed case).
:VALue?
Returns the units conversion factor.
:REFerence
[:HEIGht] <n>
Sets the head pressure height.
:HEIGht?
Returns the head pressure height.
:MODE?
Returns “OFF”, “GAS”, or LIQUID”.
:MODE OFF/GAS/LIQUID
Sets the head pressure mode.
:MEDium<n>
Sets the medium density.
:MEDium?
Returns medium.
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...