Operating Instructions - CPC8000
8.4 SCPI Error Messages and Error Codes
When an error occurs, the error icon in the lower right hand corner of the main screen turns yellow and
becomes a button. When pressed, the error log is shown and all errors are cleared. Note, going to the
error log will put the instrument in measure mode for safety reasons. Error -500 is a pneumatics error
whose message will be diferent depending on the pneumatic error. Error 999 is a general purpose error
which couldn’t be categorized in the pre-defned errors.
Table 8.4 - Error Codes SCPI
Code
Error String Returned
0
No error
-101
Undefned character
-102
Syntax error
-103
Undefned separator
-104
Parameter data type error
-109
Parameter missing
-110
Undefned header
-113
Undefned command
-114
Parameter out of range
-313
Calibration data not found
-315
Confguration data not found
-350
Errorqueue overfow
-410
Query interrupted
-500
Regulator error message
600
Default confguration not found
601
Calibration mode active! Deactivate before setting C0..C3
602
Sensor not available
701
DCS instance not available
702
Create DCS instance failed
703
DCS still active
704
Command currently not allowed
999
General purpose error message
Summary of Contents for CPC8000
Page 6: ...High End Pressure Controller CPC8000 6 Operating Instructions CPC8000 NOTES ...
Page 10: ...High End Pressure Controller CPC8000 10 Operating Instructions CPC8000 NOTES ...
Page 24: ...High End Pressure Controller CPC8000 24 Operating Instructions CPC8000 NOTES ...
Page 84: ...High End Pressure Controller CPC8000 84 Operating Instructions CPC8000 Notes ...
Page 112: ...High End Pressure Controller CPC8000 Operating Instructions CPC8000 113 NOTES ...