6.4.1.4 Tare Button
If the Tare calibration function has been chosen in the Transducer App (section 6.4.4), then the Tare But-
ton [ ] will appear in the channel screen. The Tare button and the Zero Button cannot appear on the
screen at the same time, in the same channel. When the Tare button is pressed, the instrument will sub-
tract the current pressure reading (the tare pressure) so that the indicator will display zero. Subsequent
deviations in pressure will be relative to the tare pressure. Notice in the figure below that, when active, the
back ground of the tare button is lighter than the channel background, when not active the button back-
ground is darker than the channel background.
Figure 6.4.1.4 - Tare Button
Pressing the tare button again will deactivate the tare and change the pressure indication back to the
reading corresponding to the calibrated output of the transducer. An active tare will revert to a deacti-
vated state after a power cycle.
Tare button,
active.
Tare button,
inactive.