Operating Instructions - CPG2500
55
Precision Pressure Indicator
CPG2500
Ptype?
<sp>CCCCC<cr><lf>
Returns “Absolute” or “Gauge” for the
pressure type
RangeMax?
<sp>n.nn<cr><lf>
Returns the maximum range of the ac-
tive transducer in the current units
RangeMin?
<sp>n.nn<cr><lf>
Returns the minimum range of the ac-
tive transducer in the current units
Rate?
<sp>n.nn<cr><lf>
Returns the rate reading of the instru-
ment in current units/current time unit
(see: Runits)
Rdecpt?
<sp>n<cr><lf>
Returns the number of rate decimal
points (see: Resolution)
Release?
15 char string.
Ex:
Release? Test stand 1
Returns:
<sp>(YES or NO), CCC…
CCC<cr><lf>
This command is used to release
control of the instrument in a multiple
computer environment.
Yes if release is successful
No if instrument is being controlled
with another computer
CCC… = name of controlling com-
puter or AVAILABLE
See: Acquire? and Unlock
Resolution
<n>
Sets the number of significant digits
(see: decpt)
Resolution?
<sp>n<cr><lf>
Returns the number of significant dig-
its (see: decpt)
Rfilter
Value in %
Sets the % of the rate filter
Rfilter?
<sp>n.nn<cr>
<lf>
Returns the rate filter
Runits
Sec, min, hr
Sets the rate time unit
Runits?
<sp>XXXX<cr><lf>
Returns the rate time unit
Save_cal
Save calibration values
Save_linearity
Save linearity values
Sbaud
9600, 19200, 38400, 57600, 115200 Sets the serial baud rate
Sbaud?
<sp>XXXX<cr><lf>
Returns the serial baud data
Sdata
7 or 8
Sets the serial data bits
Sdata?
<sp>n<cr><lf>
Returns the serial data bits number
Sensorid?
<sp>Mensor,SN XXXXXX, VN.NN
Returns the active sensor’s serial
number and firmware version
Span
desired pressure or ?
Sets span on active transducer or for
?, clears previous value, must be >
50% FS and has a 1% limit. CALDIS-
ABLE must be OFF/NO.
Span?
<sp>XXXXXXX<cr><lf>
Returns span scale factor for active
transducer
Sparity
Even, ODD, NONE
Sets the serial parity
Sparity?
<sp>CCCC<cr><lf>
Returns the serial parity
Summary of Contents for CPG2500
Page 14: ...14 Operating Instructions CPG2500 Precision Pressure Indicator CPG2500 NOTES ...
Page 47: ...Operating Instructions CPG2500 47 Precision Pressure Indicator CPG2500 NOTES ...
Page 61: ...Operating Instructions CPG2500 61 Precision Pressure Indicator CPG2500 NOTES ...
Page 67: ...Operating Instructions CPG2500 67 Precision Pressure Indicator CPG2500 NOTES ...
Page 69: ...Operating Instructions CPG2500 69 Precision Pressure Indicator CPG2500 NOTES ...
Page 85: ...Operating Instructions CPG2500 85 Precision Pressure Indicator CPG2500 NOTES ...