WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
Figure 6.4.1.4 Number Pad
6.4.1.4.1
Percent Step
The Percent Step entry method allows the user to select setpoints based on a percentage of the primary transducer installed.
Pressing the transducer range button at the top will open the Step Settings application (Figure 6.4.1.4.1-B). In this application, the
user can select between 2 to 12 preset steps in percentages or pressure values, overrange percent, and minimum and maximum
DUT limits. Users can also customize the setpoints by selecting the individual tiles and typing in the desired setpoints. Simply
press the transducer range button on the top right to go back to the home screen. From here, press the specific percentage or
pressure value and the setpoint will immediately change to the chosen setpoint. The up and down arrows at the bottom right of the
screen allow for nudging the pressure reading by the increments displayed to the right. These increments can be changed using
the Digital Step application.
6.4.1.4.2
Figure 6.4.1.4.1-B Step Settings
Figure 6.4.1.4.1-A Percent Step
Digit Step
The Digital Step entry method allows the user to step the setpoint up or down by increments of 1 in the selected position. For
example, in the picture below the 1 is in the tenths position, so it will step up to 10,000.1 or down to 9,999.9 upon pressing the
corresponding arrow key. The user can change the placement of the 1 by pressing the desired position signified by a white 0.
There can only be one digit representing a 1 at a time.