36
WIKA operating instructions hydraulic high pressure controller, model CPC8000-H
PN 0018708001B 11/2020 EN
EN
6. Operation
6.4.2.1.2
Auxiliary Displays
The CPC8000-H homescreen has 2 available slots for additional auxiliary displays. These two displays can be chosen on the
General settings application by selecting "Secondary Display" or "Tertiary Display". Each of these options can be configured to
show the peak, rate, sensor 3 (pilot sensor), units, tare, or uncertainty.
Figure 6.4.2.1.2 Auxiliary Displays
Some of these auxiliary displays can be modified on the main screen by pressing the button that displays the secondary or tertiary
display. Others simply display auxiliary information.
■
Peak:
This will display the upper and lower points that have been measured by the instrument. These values can be reset by
pressing the reset button on the main screen
■
Rate:
The rate display will show the changing rate of pressure in the current units. The time factor will be per second or per
minute and this can be chosen on the main screen in the auxiliary display section.
■
Sensor 3:
This option will display the pilot sensor (pneumatic pressure) in the current pressure units.
■
Units:
The units display will provide a second set of pressure units to be visible on the home screen. The units can be selected
and changed on the main screen by pressing the units auxiliary button.
■
Tare:
The tare display will show the offset/tare value if it is activated.
■
Uncertainty:
This will show the accuracy of the pressure being displayed.
6.4.2.1.3
Cal Function
The Cal Function provides the option of choosing Zero or Tare to be displayed on the main screen. Only one of the two can be
chosen at a time.
Figure 6.4.2.1.2 Cal Functions