54
WIKA operating instructions hydraulic high pressure controller, model CPC8000-H
PN 0018708001B 11/2020 EN
EN
AUXDISP?
<sp>CCCCC…<cr><lf>
Returns the secondary alternate function
mode, works identical to AUXDISP2 (kept for
backwards compatibility).
AUXDISP1
Press, peak, rate, rate setpt,
units
Sets the secondary display function mode,
works identical to AUXDISP2 (kept for
backwards compatibility).
AUXDISP1?
<sp>CCCCC…<cr><lf>
Returns the secondary alternate function
mode, works identical to AUXDISP2 (kept for
backwards compatibility).
AUXDISP2
Press, peak, rate, rate setpt,
units
Sets the secondary display function mode,
works identical to AUXDISP2 (kept for
backwards compatibility).
AUXDISP2?
<sp>CCCCC…<cr><lf>
Returns the secondary alternate function
mode, works identical to AUXDISP2 (kept for
backwards compatibility).
Baro?
<sp>+n.nn<cr><lf>
Returns reading from barometric transducer
or “NO BAROMETER” if one isn’t installed.
Baroid?
<sp>Mensor,SN XXXXXX,
VN.NN<cr><lf>
Returns identification string for the barometer.
Barocaldisable
YES,NO
Not used, kept for backwards compatibility.
Barocaldisable?
<sp>(YES or NO)<cr><lf>
Not used, kept for backwards compatibility.
Barounits
See units code or text in
Section 11.4
Sets the barometer engineering units. Note:
this only effects the displayed output on the
screen, not any remote responses.
Barounits?
<sp>CCCC<cr><lf>
Returns the barometer units in a text string
Brightness
0-100
Set the brightness of the screen from 0 to
100%
Brightness?
<sp>nn<cr><lf>
Get the brightness of the screen
Caldisable
YES,NO
Sets whether or not calibration of the active
transducer is disabled.
Caldisable?
<sp>(YES or NO)<cr><lf>
Returns whether or not calibration of the
active transducer is disabled.
Cerr
none
Clears the error queue.
Cmdset
Mensor, SCPI
Activates remote command set for instrument
emulation modes.
Cmdset?
<sp><CCCCCC><cr><lf>
Returns active command set identifier.
Control
Instrument placed in Control Mode
Control?
<sp>(YES or NO)<cr><lf>
Returns YES if instrument is in control. No if
otherwise.
Crate
Slow, Medium, Fast, Variable
Sets the control rate. Variable mode is a
predetermined user-defined rate setpt.
7. Remote Operation