57
WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
Outform
1 to 7
Sets the output format (see Section 8.2)
Outform?
<sp>X<cr><lf>
Returns the output format
Overrange
Set the overrange percentage
Overrange?
<sp>+n.nn<cr><lf>
Get the overrange percentage
Peakmax?
<sp>+n.nn<cr><lf>
Returns the maximum pressure since
peakreset was sent.
Peakmin?
<sp>+n.nn<cr><lf>
Returns the minimum pressure since
peakreset was sent.
Peakreset
None
Resets the peak values.
PILOT?
<sp>+n.nn<cr><lf>
Returns the pilot sensor (sensor 3) presssure
reading
Port
nnnnnn
Sets the Ethernet port of the instrument
Port?
<sp>nnnnn<cr><lf>
Returns the Ethernet port of the instrument
Ptype
Absolute or Gauge
Sets the instrument pressure type. Emulation
only works if the optional barometric transduc-
er is installed.
Ptype?
<sp>CCCCC<cr><lf>
Returns “GAUGE”, “GAUGE EMULATION”,
“ABSOLUTE”, or “ABSOLUTE EMULATION”.
Returns the current pressure type and notes if
it is an emulated pressure type.
Range?
<sp>+n.nn<cr><lf>
Alias for “RangeMax?”. See RangeMax?
RangeMax?
<sp>+n.nn<cr><lf>
Returns the maximum range of the active
transducer in the current units.
RangeMin?
<sp>+n.nn<cr><lf>
Returns the minimum range of the active
transducer in the current units.
Rate?
<sp>+n.nn<cr><lf>
Returns the rate reading of the instrument in
current units/current time unit.
See Runits.
Rdecpt?
<sp>n<cr><lf>
Returns the number of rate decimal points.
See Resolution
Release?
15 char string.
Ex: Release? Test_stand_1
Returns: <sp>(YES or NO),
CCC…CCC<cr><lf>
This command is used to release control of
the instrument in a multiple computer environ-
ment.
Yes if release is successful
No if instrument is being controlled with
another computer
CCC… = name of controlling computer or
AVAILABLE
See: Acquire? and Unlock
RESET_INSTRU-
MENT_TIME_ON
Reset the "instrument time on" counter to 0.
7. Remote Operation