58
WIKA operating instructions hydraulic high pressure controller, model CPC8000-H
PN 0018708001B 11/2020 EN
EN
RESET_INSTRU-
MENT_CONTROL_
TIME_ON
Reset the "instrument control time on" counter
to 0.
Resolution
<n>
Sets the number of significant digits.
See decpt
Resolution?
<sp>n<cr><lf>
Returns the number of significant digits. See
decpt
Rfilter
Value in %
Sets the % of the rate filter.
Rfilter?
<sp>+n.nn<cr><lf>
Returns the rate filter.
RFilterWin
nnn
Set the rate filter window as a floating point
value in pressure
RFilterWin?
<sp>+n.nn<cr><lf>
Returns the rate filter window
RStable?
YES or NO
RStableTime
0 to 65535
Sets the rate stable time to the number of
seconds specified
RStableTime?
<sp>XXXXXXX<cr><lf>
Returns the rate stable time
RStableWin
nnn
Alias for “RStableWindow”. See RStableWin-
dow
RStableWin?
<sp>+n.nn<cr><lf>
Alias for “RStableWindow?”. See
RStableWindow?
RStableWindow
nnn
Set the rate stable window as a %FS/second
RStableWindow?
<sp>+n.nn<cr><lf>
Returns the rate stable window
Runits
Sec, min, hr
Sets the rate time unit.
Runits?
<sp>XXXX<cr><lf>
Returns the rate time unit.
Sbaud
9600, 19200, 38400, 57600,
115200
Sets the serial baud rate.
Sbaud?
<sp>XXXX<cr><lf>
Returns the serial baud data.
Sdata
7 or 8
Sets the serial data bits.
Sdata?
<sp>n<cr><lf>
Returns the serial data bits number.
Sensor
1, 2, 3, or PRIMARY,
SECONDARY, TERTIARY
Sets the active transducer.
Sensor?
<sp>XXXXXXX,1<cr><lf>
Returns active transducer in long string
format.
Sensorid?
<sp><sen-
sor#><sp>ID<sp>MEN-
SOR,<sp><mod-
el>,<sp><SN>,<sp>V<ver-
sion><cr><lf>
Returns the active transducer’s serial number
and firmware version.
Setpoint
nnn.nnn
Sets the control setpoint for the instrument.
Value must be inside upper and lower limits.
Alias for “Setpt”
Setpoint?
<sp>nnn.nnnnE+nn<cr><lf>
Returns the control setpoint in current units
7. Remote Operation