59
WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
Setpoint%
nnn.nnn
Sets the control setpoint in %FS of primary
transducer range
Setpoint%?
<sp>nnn.nnnnE+nn<cr><lf>
Returns the current setpoint in % of primary
transducer range. Alias for “Setpt%?”
Setpt
nnn.nnn
Sets the control setpoint for the instrument.
Value must be inside upper and lower limits
Setpt?
<sp>+n.nn<cr><lf>
Returns the control setpoint in current units.
Setpt%
nnn.nnn
Sets the control setpoint in % of current range.
Setptpct
nnn.nnn
Sets the control setpoint in % of current range.
Setptpct?
<sp>+n.nn<cr><lf>
Returns the current setpoint in % of current
range.
Span
nnn.nnn (Desired pressure
value)
Sets span on active transducer or for ?, clears
previous value, must be > 50% FS and has a
1% limit. CALDISABLE must be OFF/NO.
Span?
<sp>+n.nn<cr><lf>
Returns span scale factor for active transduc-
er.
Sparity
Even, ODD, NONE
Sets the serial parity.
Sparity?
<sp>CCCC<cr><lf>
Returns the serial parity.
Srqmask
Stable,Error or both
Sets the CPC8000-H to issue a service
request (SRQ) over the IEEE when the
pressure control is stable, or an error occurs.
These are 80 hex and 40 hex respectively.
Srqmask?
<sp>{string}<cr><lf>
Returns “stable”, ”error” or “error, stable”
depending on the SRQ.
Sstop
1 or 2
Sets the serial stop bits.
Sstop?
<sp>X<cr><lf>
Returns the serial stop bits.
Stable?
Returns YES if instrument is stable, or NO.
Stabledelay
0 to 65535
Sets the stable time to the number of seconds
specified.
Stabledelay?
<sp>XXXXXXX<cr><lf>
Returns the stable time.
Stabletime
0 to 65535
Sets the stable time to the number of seconds
specified.
Stabletime?
<sp>+n.nn<cr><lf>
Returns the stable time.
StableWin
nnn.nnn
Sets the stable window as a %FS of active
transdcuer
StableWin?
<sp>+n.nn<cr><lf>
Returns the stable window.
Standby
None
Instrument placed in Standby mode
Standby?
<sp>(YES or NO)<cr><lf>
Returns YES if instrument is in Standby, NO if
otherwise
Step
Value inside upper and lower
limits and inside the range of
the active transducer
Sets the control step size for the instrument
7. Remote Operation