62
WIKA operating instructions hydraulic high pressure controller, model CPC8000-H
PN 0018708001B 11/2020 EN
EN
Table 7.7
Command or Query
Response/Function
MEASure
[:PRESsure] [R]?
Returns the pressure from range R.
:TEMPerature[R]?
Returns the temperature from range R.
:RATE[R]?
Returns the pressure rate/sec from range R.
:BAROmetric?
Returns the barometric pressure.
:SLEW[R]?
See MEASure:RATE?
CALibration
[:PRESsure][R]
:MODE?
Returns 1=calibrated or 0=not calibrated.
:DATE?
Returns date of cal “MM/DD/YY”
:DATE <i,i,i>
Sets date of cal YYYY,MM,DD
:ZERO?
Returns zero offset.
:ZERO <n>
Sets the zero offset.
:ZERO:RUN
Start autozero sequence.
:ZERO:STOP
Stop autozero sequence
:ZERO:INITiate?
Returns zero status
:ZERO:INITiate
Ignored
SENSe
[:PRESsure] [R]
:NAME?
Returns transducer name string.
:MODE?
Returns pressure type “ABSOLUTE” or “GAUGE”
:MODE ABS | GAUGE
Sets the pressure type to absolute or gauge.
:ABS?
Returns native transducer type “1” for absolute, ”0” for gauge.
:RESolution?
Returns resolution (float).
:RANGe
[:UPPer]?
Returns the maximum range.
:LOWer?
Returns the minimum range.
:Unit
[:NAME]?
Returns ASCII units (mixed case).
:VALue?
Returns the units conversion factor.
:REFerence
[:HEIGht] <n>
Sets the head pressure height.
:HEIGht?
Returns the head pressure height.
:MODE?
Returns “OFF”, “GAS”, or LIQUID”.
:MODE OFF | GAS | LIQUID
Sets the head pressure mode.
:MEDium<n>
Sets the medium density.
:MEDium?
Returns medium.
7. Remote Operation