64
WIKA operating instructions hydraulic high pressure controller, model CPC8000-H
PN 0018708001B 11/2020 EN
EN
Index
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
Unit
bar
mbar
Pa
psi
atm
kp/cm²
lbf/ft²
kPa
cmH2O(4°C)
inH2O(4°C)
inH2O(20°C)
inH2O(60°F)
ftH2O(4°C)
mmHg(0°C)
cmHg(4°C)
inHg(0°C)
inHg(60°F)
% of range
user units 1
user units 2
OUTPut
:STATe ON/OFF/1/0
ON or 1 = Control OFF or 0 = Measure.
:STATe?
Returns 0 for measure 1 for control.
:MODE MEASure/CONTrol/VENT
Sets the mode indicated.
:MODE?
Returns the mode string.
:STABle?
Returns 1 if stable 0 if not.
:AUTOvent ON/OFF/1/0
Autovent causes the controller to vent when a setpt of 0 is
sent while in gauge or gauge emulation.
:AUTOvent?
Returns the autovent state.
SOURCE
[:PRESsure]
[:LEVel]
[:IMMediate]
[:AMPLitude] <n>
Sets the setpoint.
[:AMPLitude]?
Returns the setpoint.
:SLEW <n>
Sets the slew rate setpoint in %FS/sec.
:SLEW?
Returns the slew rate setpoint.
:TOLerance <n>
Sets the stable window in %FS.
:TOLerance?
Returns the stable window in %FS.
CALCulate
:LIMit
:LOWer<n>
Sets minimum control limit.
:LOWer?
Returns minimum control limit.
:UPPer<n>
Sets maximum control limit.
7. Remote Operation