WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
7. Remote Operation
:UPPer?
Returns maximum control limit.
:VENT <n>
Pressure at which the vent solenoid opens for an uncon-
trolled vent.
:VENT?
Returns the pressure at which the vent solenoid opens.
:VRATE
Sets the rate at which the controlled vent goes to
atmosphere.
:VRATE?
Returns the rate at which the controlled vent goes to
atmosphere.
:SYSTem
:DETECT SLOW/FAST/CANCEL
Starts or stops controller adaption routine.
Note: There is no difference between slow and fast. Both
start the same controller adaptation process, and are only
for backwards compatibility.
:DETECT?
Returns status of controller adaption routine where the
response is:
0, OK
-1, active
-2, leakage or no pressure supply
-3, pressure supply too high or wrong mode
-4, default during controller adaption
-5, calculation fault
-6, abandoned process
:TARE
:STATe ON/OFF/1/0
Tare the current reading. The same tare value is applied
to all measuring transducers.
:STATe?
Returns the tare value.