20
WIKA operating instructions hydraulic high pressure controller, model CPC8000-H
PN 0018708001B 11/2020 EN
EN
5.2.1.1
CPC8000-HC
The CPC8000-HC module has 6 pressure ports: barometric reference, supply, exhaust, vent, reference, and measure/control port.
• The barometric reference port uses a barb fitting with a 10-32 unf thread. This port is used to monitor atmospheric pressure
when the optional barometric reference transducer is installed.
• The supply port uses a 7/16" SAE female thread. This port is required to have approximately 100 to 110 % of the pilot pressure
range: 35 bar.
• The exhaust port uses a 7/16" SAE female thread. This port is not used in the CPC8000-H model.
• The vent port uses a 7/16" SAE female thread. This port is used during control and vent mode to relieve pressure from the
instrument.
• The reference port uses a 7/16" SAE female thread. This port is not used in the CPC8000-H model.
• The measure/control port uses a 7/16" SAE female thread. This port is used to link the CPC8000-HC to the hydraulic module.
This provides the air supply to the intensifier while in control mode.
5.2.1.2
CPC8000-HM
The CPC8000-HM contains 3 ports: the drive air, measure/control CPC8000-HC, and measure/control DUT.
• The pneumatic supply pressure uses a male 6mm tube fitting. This port requires at least 8 bar (115 psi) of pressure, but it can
also use up to the full supply pressure of the CPC8000-HC. This pressure is used to actuate the isolation valves and priming
pump located within the hydraulic module.
• The measure/control CPC8000-HC port uses a male 6mm tube fitting. This port must be connected to the CPC8000-HC
measure/control port. It is used to supply air pressure to the intensifiers using the provided tubing.
• The measure/control DUT port uses a male 1/4" Sno-Trik fitting. Consult factory for different adapters. This port is used to
control hydraulic pressure on the DUT.
WARNING!
Do not use the instrument without the Measure/Control port connected.
Be sure to always have the Measure/Control port connected to a test sample while using the CPC8000-H. If the test
port is left open or disconnected, liquid can leak from the instrument.
Measure/Control
5. Installation
Supply Exhaust `Vent Reference
Barometric
Reference
Supply pressure
Measure/
Control from
CPC80000-HC
Measure/
Control DUT