21
WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
WARNING!
Never disconnect test ports while pressurized!
Be sure the instrument is fully vented and the source pressure is turned off while disconnecting connections and
hoses from the Measure/Control port. Failure to disassemble the test ports correctly could result in leaking liquids
and gasses.
WARNING!
Do not exceed the maximum source pressure!
The supply pressure port is limited to 110%% over the primary pilot transducer range. Do not exceed this pressure
rating labeled on the back of the instrument.
5.2.2 Communication ports
The CPC8000-H has 6 communication ports, all located on the CPC8000-HC: IEEE-488, USB device, RS-232, Ethernet, USB host
and an interface port to communicate between the two modules. See section 8, Remote Operation, for connections and
commands for operation over IEEE-488, Ethernet, USB or RS232 ports.
5.2.2.1
USB host
The USB host port is intended for future expansions or software upgrades. This port can also be accessed through the front panel.
5.2.2.2
Interface port
The interface port is a direct communication between the CPC8000-HC module and the CPC8000-HM. This provides the
feedback information from the hydraulic module transducers. The interface cable is permanently fixed to the CPC8000-HM and
can be removed from the CPC8000-HC using a slotted screwdriver.
5.2.3 Power inlet
Apply power with the included power cable to the rear power inlet port on the CPC8000-HC. This port is equiped with two 2.5 amp
250V Slo-Blo fuses and is rated at 100 - 120 VAC or 200 - 240 VAC at 50 / 60 Hz.
Figure 5.2.3 Power Inlet
DANGER!
Use caution when dealing with power cable and inlet connection. Hazard of electric shock: warning notice according
to IEC 61010-1.
Power Inlet