53
WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
7.6.4 Mensor Commands and Queries
Table 7.6.4 lists all of the current Mensor commands and queries.
Table 7.6.4
Command/Query
Data
Function/Response
?
See Section 7.2 Remote
Command Set
Returns data per the current output format.
Absolute?
<sp>YES or NO<cr><lf>
Returns if the primary transducer is native
absolute
Acquire?
15 char string.
Ex:
Acquire? Test_stand_1
Returns:
<sp>(YES or NO), CCC…
CCCcr><lf>
This command is used when multiple computers
would like to control the instrument.
Yes if acquisition is successful.
No if instrument is being controlled with
another computer.
CCC… = name of controlling computer.
See: Release? and Unlock
Address
0-31
Sets the GPIB Address.
Address?
<sp> nn <cr><lf>
Returns the GPIB Address.
All?
<sp>+n.nn,…,
+n.nn <cr><lf>
Returns pressure readings of all transducers
on the active channel, including the
barometer
Asset_tag
16 char string
General purpose string for customer use.
Asset_tag?
<sp>ssssssssssssssss<cr><lf>
Return customer asset tag string.
Autozero
none
Re-zero all the ranges. These adjustments
are not password protected and are not
saved through power cycles. This command
takes approximately 60 seconds but may
take longer depending on the time to become
stable.
Autozero?
<sp>S,T,X,X<cr><lf>
Returns autozero data where S represents
state (0 = complete, 1 = local autozero, 2 =
remote autozero), T represents the estimated
time to complete in seconds, and x is a (0)
character since this data location is not used
at this time.
Autozeroabort
none
Aborts autozero. Any transducers that have
been zeroed will not revert to previous zero
offsets.
AUX?
<sp>+n.nn<cr><lf>
Returns the auxiliary display reading.
AUXDISP
Press, peak, rate, rate setpt,
units
Sets the secondary display function mode,
works identical to AUXDISP2 (kept for
backwards compatibility).
7. Remote Operation