69
WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
Figure 8.2.3-B CPC8000-HM-M Front View
WARNING!
Check for leaks after all transducer replacements and reinstallations. Properly mount and tighten the fitting before
applying pressure to the CPC8000-H.
8.3 Recalibration
The CPC8000-H automatically adjusts the pressure reading for the effects of temperature and non-linearity within the calibrated
temperature range of 15-45C. The process is referred to as dynamic compensation because each reading is adjusted before it is
output to the display or to a communication bus. Thus, a calibrated CPC8000-H operated within its temperature band, and with
proper zero and span adjustments, will provide accurate pressure measurements.
The CPC8000-H should have the calibration verified periodically to ensure stability. The recommended calibration interval for this
instrument is given in the Specifications section under "Calibration Interval" in Section 10.1.
WARNING!
Performing calibration on components within the CPC8000-H could alter the accuracy of the complete instrument!
8.3.1 Calibration Services by Mensor or WIKA Worldwide
Mensor and WIKA worldwide have extensive experience and knowledge of Mensor products. Calibration of the transducers in
these products can be performed at the addresses below or by competent internal or external labs using the procedures in this
section.
8. Maintenance and Recalibration
Secondary
Transducer
Primary
Transducer