CLEANING & MAINTENANCE

•  REAL 8200 PLUS EL can affect and be affected by 

equipment in its immediate vicinity that generates 

electromagnetic fields; e.g alarm systems, power 

generators. Mobile phones and similar equipment 

can also interfere with the control system. Mercado 

Medic AB recommends that such equipment should 

not be stored any closer than 3.3 m(~10 ft) from the 

chair.

•  REAL 8200 PLUS EL has an electric lifting 

mechanism that can lift a person high enough to 

cause a risk of falling. If a person cannot get out 

of the chair from its highest position, then the 

seat height should never be set to higher than 

the position at which your feet touch the ground. 

The only exception to this is cases where you can 

summon help via alarm or telephone.

•   NOTE! The electrical functions on 

REAL 8200 PLUS EL chairs may be used for 

maximum 10% of the time, maximum 2 minutes 

every 20 minutes. The electronics in the chair 

measure the activity and shut down the electrical 

functions if this limit is exceeded.

•  Only electrical equipment from Mercado Medic AB 

may be used on the REAL 8200 PLUS EL.

•  Warning! Do not plug any accessory, other than 

those supplied by Mercado Medic AB, into the 

control system’s contacts.

•  The chair must be operated with care. Before 

moving off or operating the chair, consideration 

must be given to the surroundings. Any pets or 

children in the area could be easily run into. The 

chair must only be used to sit on and should not be 

regarded as a toy by either children or adults.

•  REAL 8200 PLUS EL is approved for transportation/

storage at temperatures of between -25°C and 

+60°C(-13°F and 140°F) and up to 90% non-

condensing relative humidity.

•  Max user weight,  REAL 8200 PLUS EL 150 kg 

(330 lbs).

•  The REAL 8200 PLUS EL should not be used on any 

incline greater than 5 degrees. If the chair is used or 

the brake is applied at steeper inclines, it may begin 

to slide which may cause the chair to fall over. 

•  If shipped, the chair must be transported in the 

intended cardboard box on a pallet. 

•  If the chair is transported by car, it must be strapped 

in place using straps over both the frame and the 

seat, and it should be set to the lowest possible seat 

height.

•   The chair is tested and approved for use between 

+5°C and +40°C(41°F and 104°F)  and 15 to 90%, 

non-condensing humidity. If the chair has been 

stored in an environment outside these limits, the 

chair must be left to acclimatise before use in order 

to ensure that it can be used safely.

•  Repairs and other technical measures may only be 

carried out by personnel authorised by EMP Living 

LLC.

•  Maximum height 2000m(~6500 ft) above sea level.
•  The REAL 8200 PLUS EL must not be moved when 

the seat is in the raised position.

•  Expected service-life ten (10) years.
•  REAL 8200 PLUS EL can be supplemented with a 

variety of accessories. Only accessories supplied 

by Mercado Medic should be used. All accessories 

must be fitted so that there is no risk of small parts 

coming loose. All the cables must be secured with 

cable ties to minimise the risk for strangulation.

•  Mercado Medic armrests, seats, backrests 

and other parts intended for skin contact are 

in compliance with the EU Regulation (EC) No 

1907/2006 concerning the Registration, Evaluation, 

Authorisation and Restriction of Chemicals (REACH).

IMPORTANT SAFETY INFORMATION 

CONTINUED

•  The chair should be wiped down and kept clean 

from dust and dirt. 

•  The chair should be washed using a disinfectant 

with PH 7–12 (concentrated), with the exception of 

the padding. 

•  Artificial leather has to be washed with soapy water 

or alternatively wiped off with an alcohol solution; 

e.g. disinfectant.

•  Leather upholstery requires a little extra 

maintenance;

 - Spills must be wiped up immediately with a 

damp cloth, never use strong cleaning agents.

 - Prevent upholstery from coming into contact 

with personal care products, such as make-up, 

hair gel etc.

 - Do not allow pets on leather furniture, their 

claws may scratch the leather.

 - Avoid direct and prolonged exposure to sunlight, 

it may fade the leather in the long run.

•   Seat, backrest and other upholstery are not 

manufactured in sealed materials for functional 

reasons. When reconditioning, upholstery should be 

replaced for hygienic reasons.

•   The electric lift mechanism should be checked 

regularly for dust, dirt and stability: Raise the seat 

to its highest position. Clean with a cloth. Do not use 

any water or solvents. Then lubricate the cylinder 

using a thin layer of Teflon or silicone-based grease.

Summary of Contents for REAL 8200 PLUS EL 24V SCREENING

Page 1: ...es to this manual and its contents ASSEMBLY INSTRUCTIONS INSTRUCTIONS FOR USE AND CARE REAL 8200 PLUS REAL 8200 PLUS SERIES REAL 8200 PLUS EL 24V SCREENING REAL 8200 PLUS EL 24V CLINICAL YOUR REAL 820...

Page 2: ...in the chair 5 The brake on the chair must always be applied when getting in and out of the chair The brake must be checked on a regular basis 6 Charge the chair according to the enclosed instructions...

Page 3: ...eat and it should be set to the lowest possible seat height The chair is tested and approved for use between 5 C and 40 C 41 F and 104 F and 15 to 90 non condensing humidity If the chair has been stor...

Page 4: ...sories Adjustments carried out by unauthorized service technicians Fading discoloration or natural variations occurring in skai leatherette leather and fabrics The warranty does not cover damage cause...

Page 5: ...bber part no 801264 The braking function may be reduced on uneven surfaces There are three types of brake Brake lever foot brake and electric brake 1 1 2 4 SEAT TILT Control 1 Seat tilt electric contr...

Page 6: ...on choose a wall socket that is easily accessible to plug the charger into The charger s cable is 1 8 m and must not be extended In case of emergency disconnect the chair by unplugging the charger fro...

Page 7: ...m which then shuts down one or all of the function channels and switches on a warning light on the hand control with cord The system will be reset automatically after 10 seconds If the error has been...

Page 8: ...ler in North America EMP Living LLC 14941 NE 147th CT Woodinville WA 98072 Phone 425 949 7195 Email info empliving com Web www empliving com Mercado Medic AB Phone 46 0 8 555 143 00 14 Tryffelslingan...

Reviews: