20-07316-01

7

USER MANUAL

Before using for first time

This product is delivered by Mercado Medic AB with the battery in its deactivated state. The battery must be 

activated before using for the first time. This is done by connecting the battery charger. Keep the battery charger 

connected until the battery is fully charged. This will be indicated by a green light on the charger’s LED. If you are 

using the hand control with cord, this will also be indicated by the three green battery LEDs, which will start to flash 

at the same time. The battery may also need to be charged when it has been inactive as a certain degree of self-

discharge always occurs for batteries.

Note!

Charging the battery is particularly important on chairs with electric brakes, electric seat tilting and electrical 

back functions because these functions require a high battery capacity in order to work properly. Guarantee these 

functions by recharging the battery once a day. The battery may only be changed by personnel authorised by 

Mercado Medic AB. See assembly instructions, Battery Replacement.

Troubleshooting for electrical functions on your chair

There are no movements at all:
•  The battery might need to be charged. See the instructions above.

•  An internal error may have been detected by the system, which then shuts down one or all of the function 

channels and switches on a warning light on the hand control with cord. The system will be reset automatically 

after 10 seconds. If the error has been resolved, then all channels will be unlocked and the warning light will be 

switched off.

•  The button lock may be enabled. To enable or disable the button lock for a certain function, hold in both of the 

directional arrows on the relevant button for at least three seconds.

The electric functions on the chair are slow:
•  The battery needs to be charged. See the instructions above.
If none of the above measures are able to resolve the problem, servicing may be required. Please contact your dealer 

or get in touch with Mercado Medic AB. You will find our contact details on the last page of this manual.

Emergency stop

If, for whatever reason, the system does not interrupt a movement after the corresponding function button has been 

released, the movement can be interrupted by pressing down any of the other buttons on the hand control with cord. 

If a fixed control unit is being used instead of a hand control with cord, the same emergency stop function can be 

initiated by pressing down the button for movement in another direction.

*If the chair’s battery has reached a critical level and the electronic functions do not work after six hours of charging, 

charge the chair for a further 24 hours. If the problem persists, please contact the distributor of your chair.

CHARGING CONTINUED

Summary of Contents for REAL 8200 PLUS EL 24V SCREENING

Page 1: ...es to this manual and its contents ASSEMBLY INSTRUCTIONS INSTRUCTIONS FOR USE AND CARE REAL 8200 PLUS REAL 8200 PLUS SERIES REAL 8200 PLUS EL 24V SCREENING REAL 8200 PLUS EL 24V CLINICAL YOUR REAL 820...

Page 2: ...in the chair 5 The brake on the chair must always be applied when getting in and out of the chair The brake must be checked on a regular basis 6 Charge the chair according to the enclosed instructions...

Page 3: ...eat and it should be set to the lowest possible seat height The chair is tested and approved for use between 5 C and 40 C 41 F and 104 F and 15 to 90 non condensing humidity If the chair has been stor...

Page 4: ...sories Adjustments carried out by unauthorized service technicians Fading discoloration or natural variations occurring in skai leatherette leather and fabrics The warranty does not cover damage cause...

Page 5: ...bber part no 801264 The braking function may be reduced on uneven surfaces There are three types of brake Brake lever foot brake and electric brake 1 1 2 4 SEAT TILT Control 1 Seat tilt electric contr...

Page 6: ...on choose a wall socket that is easily accessible to plug the charger into The charger s cable is 1 8 m and must not be extended In case of emergency disconnect the chair by unplugging the charger fro...

Page 7: ...m which then shuts down one or all of the function channels and switches on a warning light on the hand control with cord The system will be reset automatically after 10 seconds If the error has been...

Page 8: ...ler in North America EMP Living LLC 14941 NE 147th CT Woodinville WA 98072 Phone 425 949 7195 Email info empliving com Web www empliving com Mercado Medic AB Phone 46 0 8 555 143 00 14 Tryffelslingan...

Reviews: