RECONDITIONING

 

! Important information for reconditioning and servicing REAL 9500/9600 EL (hereinafter referred to as 

the Product):

 - The Product is not subject to requirements for preventive or periodic maintenance during the Product’s 

expected service life. However, a thorough visual inspection of the main components of the Product must 

be performed in connection with servicing and reconditioning. These include the base, lifting mechanism, 

seat frame, back recliner and U-bars for armrests. Inspections must include welds, lockable functions and 

adjustments. This is to ensure the user’s safety, as neither the Manufacturer nor the Principal has a thorough 

knowledge of the Product’s previous use.

 - The user must not sit in the Product while it is being reconditioned or serviced or during maintenance.
 - Do not use high pressure washers when cleaning the Product.
 - The serial number and delivery date are shown on the Product’s production label. The Product’s expected 

service life is ten (10) years. 

 - When checking the PLUS base, always inspect the weld between the base’s front and rear legs. Any indication 

of cracks in the material or paint in this area should be regarded as reason to replace the base, or the Product. 

 - In case of reduced braking action, check both the stability of the base and the condition of the brake rubber. 

If the Product is not resting on all four wheels in the unbraked position, the Product must be withdrawn from 

service immediately. Check for instability in different places and directions to rule out the floor as the cause of 

any instability.

 - If a Product that is more than 10 years old is to continue to be used in a safe manner, it must be fully 

refurbished. Full refurbishment must be carried out by authorised service personnel from Mercado Medic AB, 

or by reconditioning/service personnel at the Principal/Customer authorised by Mercado Medic AB. 

Full refurbishment means, among other things, replacing critical components. These include lifting mechanism, 

seat frame, back recliner and U-bars for armrests. The Product must be restored to as-new condition in such 

a way that its service life can be extended beyond that specified by the Manufacturer. A fully refurbished 

Product must comply with the requirements of the Swedish Medical Products Agency’s regulations when it is 

returned to use.  

The unit carrying out a full refurbishment takes over the role of Manufacturer and is responsible for the 

CE marking of a Product that is put back on the market. This means that if a full refurbishment has been 

performed by authorised personnel at the Principal, this unit takes over the manufacturer’s liability and 

responsibility for CE marking from Mercado Medic AB.

1. Electronics unit

 

…

1.1. Electronics/function/cleaning

Check the electronics and see which 

electrical functions are connected. 

Test electrical functions; if the chair 

does not work, try connecting the 

charger for at least 10 seconds. 

Check for any damage to the 

electronics box. 

 

…

1.2. Control unit/function/

cleaning

Check the functions of the controls, 

and check that the housing is intact 

and that all buttons are working. 

Wipe the controls with a cloth lightly 

dampened with disinfectant; this is 

to avoid the risk of cross-infection. 

Do not use solvents, bleach, polishes, 

synthetic detergents, polishing waxes 

or sprays on the chair.

 

…

1.3. Connectors, attachment

Check that cables and connectors are 

intact and properly secured. Make 

sure that cables and connectors 

cannot be pinched or crushed.

 

…

1.4. Batteries/battery charger/

function

Check that the charger is working 

and that the housing and cables are 

undamaged.
Check the batteries’ voltage 

and that there is not too great a 

difference between the batteries’ 

charge; this indicates that a battery 

cell has failed.

 

…

1.5. Electric seat tilt/ 

function

Check that the actuator is not visibly 

damaged. Check the function of the 

actuator’s moving parts. Ensure that 

all locking screws are tightened to 

avoid any play in the end position. 

 

…

1.6. Electric back/ 

function

Check that the actuator is not visibly 

damaged. Check the function of the 

actuator’s moving parts. Ensure that 

all locking screws are tightened to 

avoid any play in the end position.

 

…

1.7. Electric brake/function

Check that the actuator is not visibly 

damaged. Check the function of the 

actuator’s moving parts. Ensure that 

all locking screws are tightened to 

avoid any play in the end position.
Check all joints and that the clamp 

mounting on the brake shaft and 

base tubes is correctly positioned.

2. Electric lifting mechanism

 

…

2.1. Noise/sliding clutch

Listen for any abnormal noise in 

the electric lifting mechanism, 

and replace the electric lifting 

mechanism if a bearing is defective 

in any way. The electric lifting 

mechanism can be sent to Mercado 

Medic AB for repair/reconditioning.
Load the chair and check that the 

electric lifting mechanism does not 

slip to lowered. Otherwise ensure 

the correct torque for the sliding 

clutch.

RECONDITIONING INSTRUCTIONS CHECK POINTS

Summary of Contents for REAL 9500 EL

Page 1: ... the chair have been tested and approved in accordance with EN 1021 1 and EN 1021 2 Mercado Medic AB is certified in line with ISO 9001 ISO 14001 and ISO 13485 and complies with applicable labour and environmental legislation If you have any questions about your product or if something unexpected happens please contact your prescriber first otherwise you are welcome to contact us at Mercado Medic ...

Page 2: ......

Page 3: ...TABLE OF CONTENTS YOUR REAL 9500 9600 EL 24V PAGE 4 IMPORTANT INFORMATION CLASS PAGE 4 USER MANUAL PAGE 5 INDIVIDUAL ADAPTATION PAGE 8 CLEANING MAINTENANCE PAGE 8 SERVICE RECONDITIONING LIST PAGE 9 ...

Page 4: ...g with the brake applied on inclines of up to five degrees in all directions If the chair is parked with the brake applied on steeper inclines it may begin to slide NOTE The electrical functions on REAL 9500 9600 EL chairs may be used for max 10 of the time max 2 mins 20 mins The electronics measure the activity and shut down the electrical functions if this limit is exceeded Only tested and appro...

Page 5: ...lip locks and the back recliner cannot be pulled out The Comfort back mechanism is available in a low and high design illustrated chairs have a low back The back mechanism has separate controls for height 1 and angle of the backrest 2 Loosen the knob set the desired position for each adjustment 1 and 2 and then lock the knob The recliner angle is adjusted using the lever 3 USER MANUAL 3 4 1 Medic ...

Page 6: ...ed button 3 It has a holding circuit and only requires a short press to activate the braking function To brake the chair press button 3 again SEAT TILT The gas spring tilt is controlled by a gas spring It is operated using the lever 1 under one side of the armrest standard left The armrests are included in the tilting to ensure the best support in all tilt positions If the chair has electric tilt ...

Page 7: ...rger from the wall socket Do not sit on the chair when it is charging CONNECTING AND DISCONNECTING THE BATTERY During longer periods of storage the battery automatically goes into sleep mode after seven days The battery can also be forced into sleep mode by connecting a 9 V power supply item no 804823 to the charging outlet and waiting at least five seconds before disconnecting the power supply If...

Page 8: ... controller If a fixed control unit is being used instead of a handheld controller the same emergency stop function can be initiated by pressing the button for movement in another direction CLEANING MAINTENANCE The chair should be wiped down and kept clean from dust and dirt The chair should be washed using a disinfectant with PH 7 12 concentrated with the exception of the padding Plush upholstery...

Page 9: ...le of Manufacturer and is responsible for the CE marking of a Product that is put back on the market This means that if a full refurbishment has been performed by authorised personnel at the Principal this unit takes over the manufacturer s liability and responsibility for CE marking from Mercado Medic AB 1 Electronics unit 1 1 Electronics function cleaning Check the electronics and see which elec...

Page 10: ...ipped with crank tilt controls check that there is no play and that all screws are tightened 6 8 Coxit mechanism control Make sure that the knobs for the flaps turn easily and are securely locked with the counter nut 7 Back mechanism 7 1 Function wear Check and replace any worn parts and make sure that moving parts are working correctly 7 2 Attachment tightening Check that all screws knobs and whe...

Page 11: ......

Page 12: ... Postal address PO Box 1074 SE 181 22 Lidingö Switchboard is open Mon Fri 08 00 16 30 Tel 46 0 8 555 143 00 Fax 46 0 8 555 143 99 E mail info mercado se Space for product label with serial number Service and technical support PO Box 1074 SE 181 22 Lidingö Telephone hours for the service line Mon Fri 08 00 12 00 and 13 00 16 00 CET CEST Tel 46 0 8 555 143 08 Fax 46 0 8 555 143 99 Mobile phone 46 0 ...

Reviews: