CLEANING & CARE
• The chair should be wiped down and kept clean from dust and dirt .
• The chair should be washed using a disinfectant with PH 7–12 (concentrated), with the exception of the padding .
• Plush upholstery has to be washed with a foam wash:
1 . Remove loose dirt, crumbs and the like .
2 . Apply a thin layer of foam and rub the foam in evenly with a damp cloth .
3 . Wipe with a clean and lightly dampened cloth .
4 . Vacuum clean when dry .
•
Artificial leather has to be washed with soapy water or alternatively wiped down with an alcohol solution, e .g .
disinfectant .
• Leather upholstery requires a little extra maintenance:
- Spills must be wiped up immediately with a damp cloth; never use strong cleaning agents .
- Prevent upholstery from coming into contact with personal care products, such as make-up, hair gel, etc .
- Do not allow pets on leather furniture, as their claws may scratch the leather .
- Avoid direct and prolonged exposure to sunlight, as this can cause the leather to fade over time .
• For functional reasons, seats, backrests and other padded parts are not made from impermeable materials . Padded
parts should be replaced during reconditioning for hygienic reasons .
• The electric lifting mechanism should be checked regularly for dust, dirt and stability: Raise the seat to its highest
position . Clean with a cloth . Do not use any water or solvents . Then lubricate the cylinder using a thin layer of Teflon or
silicone-based grease .
• The user or their assistant should remove dust, wipe clean and dry off the padded parts on the chair in accordance
with the instructions above to the best of their ability and as needed . If this is not done satisfactorily, the chair will need
to be reconditioned in order to guarantee hygienic safety .
GUIDELINES FOR INDIVIDUAL ADAPTATION OF MERCADO MEDIC AB’S PRODUCTS
Adjustments may be made only by staff who have completed Mercado Medic AB's training for service and reconditioning.
Parts that are on the list of current combination agreements at
www.mercado.se/mercado-dokument
can be fitted while
retaining Mercado Medic AB’s CE marking. If a desired item that is not on the list of currentcombination agreements is
fitted, the customer’s authorised personnel must contact Mercado Medic AB for written approval.
Parts to be evaluated must be sent physically to Mercado Medic AB, either by the customer or the supplier of the part in
question. Mercado Medic AB will then contact the supplier of the part in question; if a combination agreement is deemed
to be of interest, Mercado Medic AB performs a risk analysis on the part. Once the risk analysis is complete, the customer
receives a written opinion.
Chairs that have been modified according to the following must be verified by Mercado Medic AB for Mercado Medic AB
to keep the product liability.
• Padded parts such as seats, backrests, armrests, side supports, trunk supports and neck supports can be adapted as long
as the existing external dimensions of the product are not exceeded.
• The centre of gravity over lifting devices must be maintained/centred for optimum function.
• Padded parts must be able to meet the applicable ignition requirements and other environmental requirements.
• Surfaces that are loaded with significant vertical forces must not fall outside the area (base) that is formed by the
points of the castor centres.
If these guidelines are met, Mercado Medic AB retains the product liability for its product.
Products that are specially modified by the health care services outside Mercado Medic AB’s directive, where the health
care services also take over the product liability, can be restored to their original condition. This must be carried out by an
authorised technician approved by Mercado Medic AB. Mercado Medic AB then takes back product liability in accordance
with the CE marking 93/42 EEC; Medical Devices Directive MDD.