CLEANING, MAINTENANCE AND RECHARGING
The chair must be wiped clean and kept dirt and dust
free. Clean the upholstery with foamwash; 1. Remove
dirt, crums etc. 2. Apply a thin layer of foam, use a
moist cloth to wipe the foam in an equally thick layer.
3. Swab with a clean and damp cloth. 4. Vaccum when
dry. The seat, back support and other upholstered
parts are not made from sealed materials for functional
reasons. For hygienic reasons, whenever the upholstery
is replaced, the seat and upholstery filling must also be
replaced. The electric lifting device must be checked
regularly for dirt, dust and stability: move the seat up
to the highest position. Clean with a dry cloth. Do
not use water or solvents. Lubricate the cylinder with
a thin film of teflon- or silicon-based grease. Use only
battery charger
real 9000
model 2240-12V, item no.
802289 when charging the chair. If another battery
charger is desired, contact Mercado Medic AB for a
written approval. The battery should be charged regu-
larly; if possible once a day. It is important that the
battery is never completely discharged.
CHARGING (FOR ELECTRIC CHAIR)
Connect the charger contact (
1
) into the socket on
the box where the toggle-switch for raising and lowe-
ring the seat is situated. Then plug in the charger (
2
)
to the wall socket or power point. The light on the
charger indicates when charging is taking place. The
battery is fully charged when the LED color changes
from yellow to green. Only chargers designed for
REAL-chairs may be used. The chair can be raised
and lowered while charging. This regards electric chairs
delivered from w. 11 2013 or earlier delivered upgraded
chairs.
NOTE!
Charging the battery is particularly
important for those chairs which have electric brake,
electric tilt and electric backrest, because these functions
require high battery capacity to work and to not lose
function. Avoid this by charging once a day. When
replacing battery (Model designation: HP7-12 12V
7.2Ah), make sure to that the plus battery cable is
connected to the plus pole of the battery, and the
minus cable to the minus pole. If the chair’s electric
functions stop working it is most probably due to a
blown fuse. Check the fuses in the fuse holders that are
located under the seat on the cable between the battery
and the actuator. Replace the blown fuse with a new
25A fuse.
Pay attention to the following when handling
batteries:
Never short-circuit the battery. • Don't
expose the battery to heavy bumps or open fires. If you
come into contact with battery acid, rinse with water
for 15 min. Contact a doctor. • Dead batteries must be
handed in to recycling stations. Used batteries should
be disposed at appropriate recycling facilities.
ADJUSTMENT OF THE COXIT FLAPS
ROTATION ACCESSORY (OPTIONAL)
Note!
Rotation should be in locked position when
sitting down and standing up from the chair.
Make sure that the lever has locked into place.
CONTROL
1. Turn clockwise to release and anti-clockwise to
lock the device.
REAL PLUS COXIT/EL
There are two methods for adjusting the coxit flaps.
On chairs with no seat angle adjustment the controls
are placed under the front edge of the seat. On chairs
equipped with seat angle adjustment the controls are
placed on the back edge of the seat.
CONTROL
1. Loosen, adjust to the required angle and then tighten
the knob.
2. Turn the knob to raise/lower the flap.
1
1
2
1
Charger art.no 802289
Mercado Medic AB
Telefon & Fax
Email & Internet
Tryffelslingan 14
Tel +46 8 555 143 00
info@mercado.se
181 57 Lidingö
Fax +46 8 555 143 99
Internet www.mercado.se
Sweden