CLEANING, MAINTENANCE AND RECHARGING

The chair must be wiped clean and kept dirt and dust 

free. Clean the upholstery with foamwash; 1. Remove 

dirt, crums etc. 2. Apply a thin layer of foam, use a 

moist cloth to wipe the foam in an equally thick layer. 

3. Swab with a clean and damp cloth. 4. Vaccum when 

dry. The seat, back support and other upholstered 

parts are not made from sealed materials for functional 

reasons. For hygienic reasons, whenever the upholstery 

is replaced, the seat and upholstery filling must also be 

replaced. The electric lifting device must be checked 

regularly for dirt, dust and stability: move the seat up 

to the highest position. Clean with a dry cloth. Do 

not use water or solvents. Lubricate the cylinder with 

a thin film of teflon- or silicon-based grease. Use only 

battery charger

 real 9000

 model 2240-12V, item no. 

802289 when charging the chair. If another battery 

charger is desired, contact Mercado Medic AB for a 

written approval. The battery should be charged regu-

larly; if possible once a day. It is important that the 

battery is never completely discharged.

CHARGING (FOR ELECTRIC CHAIR)

Connect the charger contact (

1

) into the socket on 

the box where the toggle-switch for raising and lowe-

ring the seat is situated. Then plug in the charger (

2

to the wall socket or power point. The light on the 

charger indicates when charging is taking place. The 

battery is fully charged when the LED color changes 

from yellow to green. Only chargers designed for 

REAL-chairs may be used.  The chair can be raised 

and lowered while charging. This regards electric chairs

delivered from w. 11 2013 or earlier delivered upgraded 

chairs. 

NOTE!

 Charging the battery is particularly 

important for those chairs which have electric brake, 

electric tilt and electric backrest, because these functions 

require high battery capacity to work and to not lose 

function. Avoid this by charging once a day. When 

replacing battery (Model designation: HP7-12 12V 

7.2Ah), make sure to that the plus battery cable is 

connected to the plus pole of the battery, and the 

minus cable to the minus pole. If the chair’s electric 

functions stop working it is most probably due to a 

blown fuse. Check the fuses in the fuse holders that are 

located under the seat on the cable between the battery 

and the actuator. Replace the blown fuse with a new 

25A fuse.

Pay attention to the following when handling 

batteries:

 Never short-circuit the battery. • Don't 

expose the battery to heavy bumps or open fires. If you 

come into contact with battery acid, rinse with water 

for 15 min. Contact a doctor. • Dead batteries must be 

handed in to recycling stations. Used batteries should 

be disposed at appropriate recycling facilities.

ADJUSTMENT OF THE COXIT FLAPS

ROTATION ACCESSORY (OPTIONAL)

Note!

 Rotation should be in locked position when 

sitting down and standing up from the chair.

Make sure that the lever has locked into place.

CONTROL

1. Turn clockwise to release and anti-clockwise to 

    lock the device. 

REAL PLUS COXIT/EL

There are two methods for adjusting the coxit flaps. 

On chairs with no seat angle adjustment the controls 

are placed under the front edge of the seat. On chairs 

equipped with seat angle adjustment the controls are 

placed on the back edge of the seat.

CONTROL

1. Loosen, adjust to the required angle and then tighten 
    the knob. 
2. Turn the knob to raise/lower the flap.

1

1

2

1

Charger art.no 802289