background image

EASYPOD

®

 HOW TO...

CHANGE THE DOSE

Dose settings

Clinical settings

Injection settings
Regional settings
Dose log
Connection

Change

Back

Remove cartridge

Menu

Device settings

Clinical settings

Teach me
Data transfer

Open

Exit

Power on the 
easypod device, 

then press the  
MENU button on  
the main screen.

Choose INPUT IN MG,  
then press OK.

Press OK to the 
warning message. 

Enter PIN code 0000 by 
using the arrow key to 
select 0 and press OK
Repeat until you see  
0000, then press OK.

Use the arrow  
key to move  

down and choose 
CLINICAL SETTINGS
then press OPEN.

Change the dose using 
the arrow buttons and 

press OK after each change.

Confirm frequency, then 
press OK. Confirm dose, 

then press OK.

Dose adjustment:  
Press OK.

If dose adjustment was OFF, after 
pressing OK you will see a screen 
asking you to confirm daily dose.
If dose adjustment was ON, after 
pressing OK you will see a dose  
spread screen with <10% or <25%
Ensure the correct value is highlighted 
in white, then press OK. Once the 
screen displays minimum and  
maximum dose spread, press OK.

Select DOSE 
SETTINGS
,  

then press CHANGE.

Always check that the dose is correct at the top of the easypod main screen.

1

5

2

6

7

8

3

4

The dose has now been 
updated and you are back  
in the Clinical Settings menu.
If you do not need to make 
any other changes, press 
BACK again, then press EXIT 
to return to the main screen.