Cellicon
TM
Perfusion Solution for Process Development User Guide
6
Tapping the icons listed in the following table displays the sensor
output:
Icon
Display
Pump Speed in RPM
F1
Feed flow rate in mL/min
P1
Feed pressure in selected pressure units
P2
Retentate pressure in selected pressure units
P3
Perfusate (permeate) pressure in selected pressure units
F2
Perfusate (permeate) flow rate in mL/min
The icons change color to indicate the alert state of each component:
Color
Alert Type
Red
1
Yellow
2
Blue
3
for details on the alert classifications.
Control
Change the pump state, setpoint, zero-adjust and access guidance for
priming from this screen.
Flow
Setpoint
Prime
System
Zero
Adjust
Flow Rate
Speed
Setpoint
Pump
Start
Decrease
Flow Rate
Increase
Flow Rate
Status
The current values, state of the system components and data logging
switch are displayed here.
TMP Indicator
Data Logging
Control
Component Indicators
Data Logging
Control
The TMP is a calculation based on P1, P2 and P3. It will only display a
value when all three pressure sensors are connected.
The data logging feature requires a USB drive. See
details about how to use data logging.
Each box within the status bar change color to indicate the status of
each component:
Color
Component Status
Grey
Not connected
White
Connected and ready to use (except pump, which
will display a type 1 alert until single-use pump
head is loaded)
Blue
Connected with a Type 3 alert
Yellow
Connected with a Type 2 alert
Red
Connected with a Type 1 alert